EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

DESARROLLO DE UN SIMULADOR FÍSICO PARA COMPARAR DOSIS ABSORBIDA EN UNIDADES DE BRAQUITERAPIA DE MEDIANA Y ALTA TASA DE DOSIS

Luis Rangel Piñero, Carlos Rodríguez Montero

Perfiles · 23.11.2026 · 1 (36) · 38-52

DOI: 10.47187/perf.v1i36.410 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Perfiles, online erschienen am 23.11.2026. Verfasst von Luis Rangel Piñero und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This is a dosimetric intercomparison study aimed at strengthening quality control in brachytherapy. The primary objective was to evaluate the agreement between the planned dose and the actual absorbed dose in clinical settings using different irradiation technologies. Methodology: A custom-designed polymethyl methacrylate (PMMA) phantom was used for the simultaneous insertion of: a Farmer NE2571 camera, a TLD-100, EBT2 films, and PTW diodes (9192/9193). The study compared treatments using 137Cs (MDR) and 60Co (HDR) sources with a standard treatment plan of 200 cGy, analyzing the impact of calculation algorithms and the quality of diagnostic images.

Results

Significant differences were found between the planning systems and the experimental measurements. In institutions that use 137Cs, deviations of up to 51.2% were recorded, attributed to geometric uncertainty due to the obsolescence of X-ray equipment and limitations in the densitometric reading of films. The diodes enabled effective verification, detecting underestimations linked to radiation-induced sensitivity degradation.

Conclusions

The accuracy of brachytherapy depends on image quality and the use of up-to-date planning algorithms. It is imperative to standardize external audit protocols to ensure radiation safety and treatment efficacy.

Zitieren

APA

Piñero, L. R., & Montero, C. R. (2026). DESARROLLO DE UN SIMULADOR FÍSICO PARA COMPARAR DOSIS ABSORBIDA EN UNIDADES DE BRAQUITERAPIA DE MEDIANA Y ALTA TASA DE DOSIS. Perfiles, 1(36), 38-52. https://doi.org/10.47187/perf.v1i36.410

Vancouver

Piñero LR, Montero CR. DESARROLLO DE UN SIMULADOR FÍSICO PARA COMPARAR DOSIS ABSORBIDA EN UNIDADES DE BRAQUITERAPIA DE MEDIANA Y ALTA TASA DE DOSIS. Perfiles. 2026;1(36):38-52. doi:10.47187/perf.v1i36.410

BibTeX
@article{Piero2026,
  author = {Piñero, Luis Rangel and Montero, Carlos Rodríguez},
  title = {DESARROLLO DE UN SIMULADOR FÍSICO PARA COMPARAR DOSIS ABSORBIDA EN UNIDADES DE BRAQUITERAPIA DE MEDIANA Y ALTA TASA DE DOSIS},
  journal = {Perfiles},
  year = {2026},
  volume = {1},
  number = {36},
  pages = {38--52},
  doi = {10.47187/perf.v1i36.410}
}
Kostenlos starten →