Peripheral Corneal Burns and Presumed Limbal Stem Cell Deficiency Following Transscleral Diode Cyclophotocoagulation: A Case Report
Journal of Ophthalmology and Advance Research · 29.06.2026 · 7 (2) · 1-5
DOI: 10.46889/joar.2026.7212 ↗Kurz gefasst
Fallbericht in der Zeitschrift Journal of Ophthalmology and Advance Research, online erschienen am 29.06.2026. Verfasst von Saleena Nurmohamed und 3 weiteren Autor:innen. Die Arbeit nennt 11 Literaturangaben.
Abstract (Originalsprache)
Background
Transscleral diode cyclophotocoagulation (cyclodiode laser) is an established treatment for refractory glaucoma which reduces aqueous humour production through targeted ablation of the ciliary body, thus reducing intraocular pressure. Though considered relatively safe, inappropriate probe positioning or excessive anterior treatment may cause thermal injury to adjacent ocular structures. Corneal burns and limbal stem cell deficiency are uncommon but potentially sight-threatening complications. Case Presentation: A 78 year old male with left neovascular glaucoma on a background of proliferative diabetic retinopathy and rubeosis iridis underwent left eye inferior 180° cyclodiode for poorly controlled intraocular pressure despite maximal topical therapy. The laser was set to a power of 1500 to 2000 milliwatts and duration of 1500 to 2000 milliseconds with eighteen applications. The postoperative note highlighted burns on some laser spots with subsequent reduction in the duration and power. Within days of receiving cyclodiode the patient developed severe ocular pain, photophobia, foreign body sensation and keratitis. There was a progressive reduction in visual acuity in the left eye from 6/18 preoperatively to hand movements vision in two months. There was associated corneoscleral thinning, severe corneal burns and ocular surface instability which was clinically consistent with iatrogenic limbal stem cell deficiency secondary to thermal injury.
Conclusion
This case demonstrates a rare but serious anterior segment complication following cyclodiode. It highlights the importance of meticulous technique, correct probe positioning, careful energy titration and awareness of limbal anatomy to minimise the risk of anterior segment injury. Early recognition and management of complications may improve long term visual outcomes.
Zitieren
APA
Nurmohamed, S., Sharma, O., Khan, H., & Chakrabarti, M. (2026). Peripheral Corneal Burns and Presumed Limbal Stem Cell Deficiency Following Transscleral Diode Cyclophotocoagulation: A Case Report. Journal of Ophthalmology and Advance Research, 7(2), 1-5. https://doi.org/10.46889/joar.2026.7212
Vancouver
Nurmohamed S, Sharma O, Khan H, Chakrabarti M. Peripheral Corneal Burns and Presumed Limbal Stem Cell Deficiency Following Transscleral Diode Cyclophotocoagulation: A Case Report. Journal of Ophthalmology and Advance Research. 2026;7(2):1-5. doi:10.46889/joar.2026.7212
BibTeX
@article{Nurmohamed2026,
author = {Nurmohamed, Saleena and Sharma, Ojasvi and Khan, Habib and Chakrabarti, Monali},
title = {Peripheral Corneal Burns and Presumed Limbal Stem Cell Deficiency Following Transscleral Diode Cyclophotocoagulation: A Case Report},
journal = {Journal of Ophthalmology and Advance Research},
year = {2026},
volume = {7},
number = {2},
pages = {1--5},
doi = {10.46889/joar.2026.7212}
}