Six Sigma-based assessment of critical cardiac biomarkers in the emergency laboratory

Elif Bengu Gungor Ay, Fatma Ucar, Aziz Sener

Annals of Clinical and Analytical Medicine · 01.11.2026 · 17 (11) · 1177-1181

DOI: 10.4328/acam.50032 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Annals of Clinical and Analytical Medicine, erschienen am 01.11.2026. Verfasst von Elif Bengu Gungor Ay und 2 weiteren Autor:innen. Die Arbeit nennt 24 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aim

Emergency laboratories require rapid, reliable test results to ensure patient safety, particularly for critical cardiac parameters. Six Sigma methodology offers an objective approach to assessing analytical performance by integrating precision, bias, and total allowable error. This study aimed to evaluate the analytical performance of high-sensitivity troponin T (Hs-TnT), creatine kinase-MB (CK-MB), and N-terminal pro–B-type natriuretic peptide (NT-proBNP) utilizing sigma metrics on two automated analyzers.

Methods

This study was conducted in the Clinical Biochemistry Laboratory of Ankara Etlik City Hospital using internal quality control (IQC) and external quality control (EQC) data collected between July and December 2024. Sigma metrics were calculated for Hs-TnT, CK-MB, and NT-proBNP on two Roche Cobas 8000 e801 analyzers. CV%, bias%, and TEa values (Clinical Laboratory Improvement Amendments [CLIA] 2025) were used to compute sigma. Quality goal index (QGI) was used to further explore sources of analytical error when sigma values were below 3.

Results

Hs-TnT demonstrated excellent performance (σ ≥ 6) on both analyzers at Level 2, while NT-proBNP showed excellent performance on Analyzer A at Level 1. No analyte exhibited poor performance (σ < 3) on either analyzer. CK-MB displayed adequate to good performance depending on the analyzer and control level. QGI results supported that none of the biomarkers showed error patterns indicative of critical analytical problems.

Conclusion

Hs-TnT, NT-proBNP, and CK-MB demonstrated acceptable-to-excellent analytical performance according to Sigma metrics, confirming their reliability in emergency laboratory settings. Sigma methodology is a fundamental approach for evaluating analytical performance, especially for critical tests. Ongoing assessment and verification ensure analytical quality and reliable results for clinical decision-making.

Zitieren

APA

Ay, E. B. G., Ucar, F., & Sener, A. (2026). Six Sigma-based assessment of critical cardiac biomarkers in the emergency laboratory. Annals of Clinical and Analytical Medicine, 17(11), 1177-1181. https://doi.org/10.4328/acam.50032

Vancouver

Ay EBG, Ucar F, Sener A. Six Sigma-based assessment of critical cardiac biomarkers in the emergency laboratory. Annals of Clinical and Analytical Medicine. 2026;17(11):1177-1181. doi:10.4328/acam.50032

BibTeX
@article{Ay2026,
  author = {Ay, Elif Bengu Gungor and Ucar, Fatma and Sener, Aziz},
  title = {Six Sigma-based assessment of critical cardiac biomarkers in the emergency laboratory},
  journal = {Annals of Clinical and Analytical Medicine},
  year = {2026},
  volume = {17},
  number = {11},
  pages = {1177--1181},
  doi = {10.4328/acam.50032}
}
Kostenlos starten →