Diacare point-of-care versus microhematocrit manual methods for determination of haematocrit levels in pregnant women at booking visit in a Tertiary Institution South Eastern Nigeria.

KE Ezenwa, CP Nnamani, O Agada, O Okorie

Nigerian Frontline Medicine Journal · 2026 · 4 (1) · 14-22

DOI: 10.4314/nfmj.v4i1.3 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Nigerian Frontline Medicine Journal, 2026. Verfasst von KE Ezenwa und 3 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Anaemia in pregnancy is a global health problem especially in resource poor countries like Nigeria. Point of care test such as Diacare EasyHB haemoglobin meter is a dual haemoglobin/haematocrit point of care device which is cheap, easy, convenient to perform and is of great value in low resource settings where availability of personnel expertise and electricity may be lacking. To establish whether the haematocrit levels determined from Diacare Easy HB haemoglobin meter among pregnant women at booking visit was comparable to the standard manual microhaematocrit method.

Methods

A cross-sectional hospital-based study carried out between February to May 2023. A total of 294 pregnant women were recruited at booking visit. Their haematocrit levels were determined using both the Diacare Easy HB Haemoglobin Meter POCT and manual microhaematocrit methods.

Results

The mean haematocrit level from the Diacare Easy HB Haemoglobin meter POCT was 31.01+2.78% while, the mean haematocrit levels from the manual microhaematocrit method were 31.60+2.66%. The mean difference of both PCV levels was 0.59 (95% CI; 0.37-0.81). The accuracy of the Diacare EasyHB haemoglobin meter POCT was 0.57 which was limited discriminatory performance and significant (p=0.024). The Pearson's correlation of the Diacare EasyHB haemoglobin meter POCT to the manual microhaematocrit showed a strong, significant positive linear relationship (0.77; p=0.001). The agreement between both methods was moderate (0.52) and significant p=0.001. The sensitivity of the Diacare EasyHB haemoglobin meter POCT from the Kappa test was 86.6%. The specificity of the Diacare EasyHB haemoglobin meter POCT from the Kappa test was 75.0%.

Conclusion

This study showed that the Diacare EasyHB haemoglobin meter method had moderate accuracy in determination of haematocrit levels among pregnant women at booking visit.

Zitieren

APA

Ezenwa, K., Nnamani, C., Agada, O., & Okorie, O. (2026). Diacare point-of-care versus microhematocrit manual methods for determination of haematocrit levels in pregnant women at booking visit in a Tertiary Institution South Eastern Nigeria. Nigerian Frontline Medicine Journal, 4(1), 14-22. https://doi.org/10.4314/nfmj.v4i1.3

Vancouver

Ezenwa K, Nnamani C, Agada O, Okorie O. Diacare point-of-care versus microhematocrit manual methods for determination of haematocrit levels in pregnant women at booking visit in a Tertiary Institution South Eastern Nigeria. Nigerian Frontline Medicine Journal. 2026;4(1):14-22. doi:10.4314/nfmj.v4i1.3

BibTeX
@article{Ezenwa2026,
  author = {Ezenwa, KE and Nnamani, CP and Agada, O and Okorie, O},
  title = {Diacare point-of-care versus microhematocrit manual methods for determination of haematocrit levels in pregnant women at booking visit in a Tertiary Institution South Eastern Nigeria},
  journal = {Nigerian Frontline Medicine Journal},
  year = {2026},
  volume = {4},
  number = {1},
  pages = {14--22},
  doi = {10.4314/nfmj.v4i1.3}
}
Kostenlos starten →