Stapler Malfunctions in Bariatric Surgery: An Analysis of the MAUDE Database
JSLS : Journal of the Society of Laparoscopic & Robotic Surgeons · 2022 · 26 (1) · e2021.00074
DOI: 10.4293/jsls.2021.00074 ↗Kurz gefasst
Fachartikel in der Zeitschrift JSLS : Journal of the Society of Laparoscopic & Robotic Surgeons, 2022. Verfasst von Benjamin Clapp und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery, Texas Tech HSC Paul Foster School of Medicine, El Paso, TX, USA. Hauptthemen laut MEDLINE: Gastric Bypass. Die Arbeit wurde bisher 16-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Staple line leaks are a serious problem in bariatric surgery and a major cause of serious morbidity and mortality. Adverse events caused by medical devices are reported to the Food and Drug Administration which maintains the Manufacturer and User Facility Device Experience (MAUDE) database. We examined adverse stapler events reported to the MAUDE database, specifically with regards to bariatric surgery.
Methods
The MAUDE database was queried for adverse events caused by staplers between January 1, 2018 - December 31, 2020; events reported by Intuitive, Ethicon, and Medtronic/Covidien; and limited our search to "gastric bypass", "sleeve gastrectomy", "stapler malfunction" combined with each company.
Results
There were 883 adverse events reported for Medtronic, 353 for Ethicon, and 35 for Intuitive. Approximately 3.5 million staple reloads sold in the study period. The reported misfire rate for Medtronic was 0.04% and for Ethicon was 0.02%. Data for Intuitive was unavailable. The most common reported event for Medtronic was failure to fire (n = 349), followed by misfire (n = 186). For Ethicon, the most common event was failure to fire (n = 146), followed by mechanical problems (n = 27). The most common event with the Intuitive stapler was leak (n = 10) and bleeding from staple line (n = 8).
Conclusions
Stapler malfunction is a very rare event in metabolic and bariatric surgery. All of the major stapler producers have transitioned to powered staplers with excellent safety profiles. Open and honest reporting about stapler malfunction is essential to determine the true safety of these ubiquitous devices.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Surgery, Texas Tech HSC Paul Foster School of Medicine, El Paso, TX, USA.
- Universidad Autonoma Guadalajara, Guadalajara, Mexico.
Zitieren
APA
Clapp, B., Schrodt, A., Ahmad, M., Wicker, E., Sharma, N., Vivar, A., & Davis, B. (2022). Stapler Malfunctions in Bariatric Surgery: An Analysis of the MAUDE Database. JSLS : Journal of the Society of Laparoscopic & Robotic Surgeons, 26(1), e2021.00074. https://doi.org/10.4293/jsls.2021.00074
Vancouver
Clapp B, Schrodt A, Ahmad M, Wicker E, Sharma N, Vivar A, et al. Stapler Malfunctions in Bariatric Surgery: An Analysis of the MAUDE Database. JSLS : Journal of the Society of Laparoscopic & Robotic Surgeons. 2022;26(1):e2021.00074. doi:10.4293/jsls.2021.00074
BibTeX
@article{Clapp2022,
author = {Clapp, Benjamin and Schrodt, Alexander and Ahmad, Maria and Wicker, Ellen and Sharma, Nishtha and Vivar, Andres and Davis, Brian},
title = {Stapler Malfunctions in Bariatric Surgery: An Analysis of the MAUDE Database},
journal = {JSLS : Journal of the Society of Laparoscopic & Robotic Surgeons},
year = {2022},
volume = {26},
number = {1},
pages = {e2021.00074},
doi = {10.4293/jsls.2021.00074}
}