EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Conformity Assessment of Lead Aprons Used in Conventional Radiology: A Multi-Centre Survey in Togo

Massaga Dagbé, Bidamin N’timon, Ousmane Aminata Bah, Pihou Gbandé, Komi Awobanou, Koku Adambounou, Masamaesso Tchaou, Abdoulatif Amadou, Lantam Sonhaye, Lama Kegdigoma Agoda-koussema, Komlanvi Victor Adjénou

Open Journal of Radiology · 2024 · 14 (02) · 55-63

DOI: 10.4236/ojrad.2024.142006 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Open Journal of Radiology, 2024. Verfasst von Massaga Dagbé und 10 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aim

Lead aprons are used to protect against scattered radiation from the patient during interventional procedures and certain special conventional radiological examinations. Given the importance of the role lead aprons are supposed to play in radiation protection, we propose to assess their conformity in medical imaging departments in public and religious hospitals in Togo. Materials and method: A multi-centre survey conducted from 26 November to 06 December 2021 in the radiology departments of public and religious health facilities in Togo. All aprons in use were included. The evaluation criteria were physical (visual), quantitative (radiographic) and qualitative (dosimetric).

Results

We had registred 43 aprons among wich 27 (62.79%) leaded aprons were labelled non-compliant and 16 (37.21%) were labelled compliant. Of the aprons judged to be non-compliant, 70.37% were more than 10 years old and 96.30% showed defects on the radiographic images. The most common defects were vampire marks (18.64%), multiple folds (16.96%), cracks (16.96%), multiple cracks (15.25%), tears (8.47%), absence of lead (5.08%), holes (3.39%) and lead corrosion (1.69%). Defective aprons (62.96%) had at least two defects. The defects were of thoraco-abdomino-pelvic (74.07%), thoracic (14.82%) and abdomino-pelvic (11.11%) topography. For indirect exposure at 50 and 70 kilovolts, all the aprons had an attenuation factor greater than 90%. After dosimetric measurement, 13.95% of aprons had attenuation factors below 90% for indirect exposure at 100 kilovolts.

Conclusion

The compliance of the leaded decks is trifactorial (physical, radiographic and dosimetric). However, there is no significant difference in X-ray attenuation capacity between defective and normal decks.

Zitieren

APA

Dagbé, M., N’timon, B., Bah, O. A., Gbandé, P., Awobanou, K., Adambounou, K., Tchaou, M., Amadou, A., Sonhaye, L., Agoda-koussema, L. K., & Adjénou, K. V. (2024). Conformity Assessment of Lead Aprons Used in Conventional Radiology: A Multi-Centre Survey in Togo. Open Journal of Radiology, 14(02), 55-63. https://doi.org/10.4236/ojrad.2024.142006

Vancouver

Dagbé M, N’timon B, Bah OA, Gbandé P, Awobanou K, Adambounou K, et al. Conformity Assessment of Lead Aprons Used in Conventional Radiology: A Multi-Centre Survey in Togo. Open Journal of Radiology. 2024;14(02):55-63. doi:10.4236/ojrad.2024.142006

BibTeX
@article{Dagb2024,
  author = {Dagbé, Massaga and N’timon, Bidamin and Bah, Ousmane Aminata and Gbandé, Pihou and Awobanou, Komi and Adambounou, Koku and Tchaou, Masamaesso and Amadou, Abdoulatif and Sonhaye, Lantam and Agoda-koussema, Lama Kegdigoma and Adjénou, Komlanvi Victor},
  title = {Conformity Assessment of Lead Aprons Used in Conventional Radiology: A Multi-Centre Survey in Togo},
  journal = {Open Journal of Radiology},
  year = {2024},
  volume = {14},
  number = {02},
  pages = {55--63},
  doi = {10.4236/ojrad.2024.142006}
}
Kostenlos starten →