The Accuracy of Readily Available Consumer-Grade Oxygen Saturation Monitors in Pediatric Patients
Respiratory Care · 2024-04 · 69 (4) · 387-394
DOI: 10.4187/respcare.11258 ↗Kurz gefasst
Fachartikel in der Zeitschrift Respiratory Care, online erschienen am 27.03.2024. Verfasst von Thomas Kovesi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Children's Hospital of Eastern Ontario, University of Ottawa, Ottawa, Canada. kovesi@cheo.on.ca. Hauptthemen laut MEDLINE: Oxygen Saturation, Mobile Applications. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
BACKGROUND: Pulse oximetry measurement is ubiquitous in acute health care settings in high-income countries and is familiar to any parent whose child has been treated in such a setting. Oximeters for home use are readily available online and are incorporated in several smartphones and smartwatches. METHODS: We wished to determine how accurate are oximeters available online that are designated for adult and pediatric use, and the saturation monitor integrated in a smartphone, when used in children, compared to reference, hospital-grade oximeters. We evaluated a fingertip oximeter marketed for children purchased online; an adult fingertip oximeter purchased online; the oximeter integrated in a smartphone; and reference, hospital-grade oximeters. Participants were < 18 y of age. Bland-Altman charts were generated, and the estimated root mean square error (EA RMS ) was calculated. Rates of failure to obtain a measurement, relationship between device and time to successful measurement, relationship between age and time to successful measurement, and relationship between error (vs the reference device) and age were evaluated for each consumer-grade device. RESULTS: We measured S pO 2 in 74 children between 0.1–17.0 y of age. Subjects weighing < 30 kg had a median (interquartile range [IQR]) age of 2 (1.0 month–1.4 y) months, and subjects weighing ≥ 30 kg had a median (IQR) age of 14.3 (11.9–16.2) y. Readings could not be obtained in 7.5, 0, and 38.8% of subjects using the pediatric, adult, and smartphone oximeters, respectively. The time to successful reading had a modest negative correlation with age with the inexpensive adult and pediatric oximeters. The inexpensive pediatric oximeter had an overall negative bias, with a mean difference from the reference device of −4.5% (SD 7.9%) and an error that ranged from > 8% to < 33% the reference device. The EA RMS was 7.92%. The inexpensive adult oximeter demonstrated no obvious trend in error in the limited saturation range evaluated of 87–99%. The overall mean difference was −0.7% (SD 2.5%). EA RMS was 2.5%. The smartphone oximeter underestimated S pO 2 at saturations < 94% and overestimated S pO 2 for saturations > 94%. Saturations could read as much as > 4%, or < 17%, than the reference oximeter. The mean difference was −2.9% (SD 5.2%). EA RMS was 5.1%. CONCLUSIONS: Our findings suggest that the performance of consumer-grade devices varies considerably by both subject age and device. The pediatric fingertip device and smartphone application we tested are poorly suited for use in infants. The adult fingertip device we tested performed quite well in larger children with relatively normal oxygen saturations, and the pediatric fingertip device performed moderately well in subjects > 1 y of age who weighed < 30 kg. Given the vast number of devices available online and ever-changing technology, research to evaluate nonclinical oximeters will continue to be required.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Children's Hospital of Eastern Ontario, University of Ottawa, Ottawa, Canada. kovesi@cheo.on.ca.
- Children's Hospital of Eastern Ontario, University of Ottawa, Ottawa, Canada.
- University of Calgary, Calgary, Canada.
- Children's Hospital of Eastern Ontario Research Institute, University of Ottawa, Ottawa, Canada.
Zitieren
APA
Kovesi, T., Saban, J., Haddad, J. F., Reddy, D., Webster, R., & Udupa, S. (2024). The Accuracy of Readily Available Consumer-Grade Oxygen Saturation Monitors in Pediatric Patients. Respiratory Care, 69(4), 387-394. https://doi.org/10.4187/respcare.11258
Vancouver
Kovesi T, Saban J, Haddad JF, Reddy D, Webster R, Udupa S. The Accuracy of Readily Available Consumer-Grade Oxygen Saturation Monitors in Pediatric Patients. Respiratory Care. 2024;69(4):387-394. doi:10.4187/respcare.11258
BibTeX
@article{Kovesi2024,
author = {Kovesi, Thomas and Saban, Jeremy and Haddad, John Faissal and Reddy, Deepti and Webster, Richard and Udupa, Sharmila},
title = {The Accuracy of Readily Available Consumer-Grade Oxygen Saturation Monitors in Pediatric Patients},
journal = {Respiratory Care},
year = {2024},
volume = {69},
number = {4},
pages = {387--394},
doi = {10.4187/respcare.11258}
}
EUVIMED