The Precision Between Transcutaneous Carbon Dioxide Versus P aCO 2 in Infants Undergoing Therapeutic Hypothermia

Pranav Garlapati, Payam Vali, Satyan Lakshminrusimha, Brian J Smith, Gerald S Zavorsky

Respiratory Care · 2024

DOI: 10.4187/respcare.10696 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Respiratory Care, 2024. Verfasst von Pranav Garlapati und 4 weiteren Autor:innen. Beteiligte Einrichtung: Dr Garlapati is affiliated with Newborn Specialists of Tulsa, Tulsa, Oklahoma. Drs Vali and Lakshminrusimha are affiliated with Department of Pediatr…. Hauptthemen laut MEDLINE: Carbon Dioxide, Hypothermia, Induced. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Infants with hypoxic-ischemic encephalopathy are often treated with therapeutic hypothermia and high-frequency ventilation. Fluctuations in PaCO2 during therapeutic hypothermia are associated with poor neurodevelopmental outcomes. Transcutaneous CO2 monitors offer a noninvasive estimate of PaCO2 represented by transcutaneously measured partial pressure of carbon dioxide (PtcCO2 ). We aimed to assess the precision between PtcCO2 and PaCO2 values in neonates undergoing therapeutic hypothermia. METHODS: This was a retrospective chart review of 10 neonates who underwent therapeutic hypothermia requiring respiratory support over 2 y. A range of 2-27 simultaneous PtcCO2 and PaCO2 pairs of measurements per neonate were analyzed via linear mixed models and a Bland-Altman plot for multiple observations per neonate. RESULTS: A linear mixed-effect model demonstrated that PtcCO2 and PaCO2 (controlling for sex) were similar. The 95% CI of the mean difference ranged from -2.3 to 5.7 mm Hg (P = .41). However, precision was poor as the PtcCO2 ranged from > 18 mm Hg to < 13 mm Hg than PaCO2 values for 95% of observations. CONCLUSIONS: The neonates' PtcCO2 was as much as 18 mm Hg higher to 13 mm Hg lower than the PaCO2 95% of the time. Transcutaneous CO2 monitoring may not be a good trending tool, nor is it appropriate for estimating PaCO2 in patients undergoing therapeutic hypothermia.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Carbon DioxideHypothermia, InducedInfant, NewbornHumansBlood Gas Monitoring, TranscutaneousRetrospective StudiesPartial Pressure

Stichwörter der Autor:innen

blood gas analysisblood gas monitoringhypothermia inducedinfantnewbornstatisticstranscutaneous

Substanzen

Carbon Dioxide

Autor:innen und Institutionen

  • Dr Garlapati is affiliated with Newborn Specialists of Tulsa, Tulsa, Oklahoma. Drs Vali and Lakshminrusimha are affiliated with Department of Pediatrics, UC Davis Medical Center, Sacramento, California. Mr Smith is affi…

Zitieren

APA

Garlapati, P., Vali, P., Lakshminrusimha, S., Smith, B. J., & Zavorsky, G. S. (2024). The Precision Between Transcutaneous Carbon Dioxide Versus P aCO 2 in Infants Undergoing Therapeutic Hypothermia. Respiratory Care. https://doi.org/10.4187/respcare.10696

Vancouver

Garlapati P, Vali P, Lakshminrusimha S, Smith BJ, Zavorsky GS. The Precision Between Transcutaneous Carbon Dioxide Versus P aCO 2 in Infants Undergoing Therapeutic Hypothermia. Respiratory Care. 2024. doi:10.4187/respcare.10696

BibTeX
@article{Garlapati2024,
  author = {Garlapati, Pranav and Vali, Payam and Lakshminrusimha, Satyan and Smith, Brian J and Zavorsky, Gerald S},
  title = {The Precision Between Transcutaneous Carbon Dioxide Versus P aCO 2 in Infants Undergoing Therapeutic Hypothermia},
  journal = {Respiratory Care},
  year = {2024},
  doi = {10.4187/respcare.10696}
}
Kostenlos starten →