Development, Validation and Application of an Ion-Pair Reversed-Phase Liquid Chromatography-Tandem Mass Spectrometry Method for the Quantification of Nusinersen
Bioanalysis · 2024
DOI: 10.4155/bio-2023-0218 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioanalysis, 2024. Verfasst von Xiao Zhang und 6 weiteren Autor:innen. Beteiligte Einrichtung: School of Pharmacy, Key Laboratory of Molecular Pharmacology & Drug Evaluation (Yantai University), Yantai University, Yantai, 264005, China. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, Chromatography, Reverse-Phase, Oligonucleotides. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 32 Literaturangaben.
Abstract (Originalsprache)
Background: The fully phosphorothioate-modified oligonucleotide (OGN) nusinersen has low ionization efficiency in the negative ion mode, resulting in a low mass spectrometry response. There have been no relevant reports on developing a LC-MS method for the determination of nusinersen by optimizing mobile phase composition. Materials & methods: Mobile phase additives comprised of 15 mM triethylamine/25 mM 1,1,1,3,3,3-hexafluoro-2-propanol with a pH of 9.6. Nusinersen was extracted from plasma using Oasis® HLB solid-phase extraction (Waters, MA, USA). Results & conclusion: By adjusting the pH of the mobile phase to 9.6 by optimizing the type and concentration of ion-pair reagents, a high mass spectrometry response was obtained. The developed method was applied to nusinersen and met the requirements for the pharmacokinetic study of nusinersen in rabbits.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Pharmacy, Key Laboratory of Molecular Pharmacology & Drug Evaluation (Yantai University), Yantai University, Yantai, 264005, China.
- Department of Traditional Chinese Medicine, Shandong College of Traditional Chinese Medicine, Yantai, China.
- National Key Laboratory of Advanced Drug Delivery & Release Systems, Luye Pharmaceutical Group, Yantai, China.
Zitieren
APA
Zhang, X., Sha, C., Zhang, W., Zhao, F., Zhu, M., Leng, G., & Liu, W. (2024). Development, Validation and Application of an Ion-Pair Reversed-Phase Liquid Chromatography-Tandem Mass Spectrometry Method for the Quantification of Nusinersen. Bioanalysis. https://doi.org/10.4155/bio-2023-0218
Vancouver
Zhang X, Sha C, Zhang W, Zhao F, Zhu M, Leng G, et al. Development, Validation and Application of an Ion-Pair Reversed-Phase Liquid Chromatography-Tandem Mass Spectrometry Method for the Quantification of Nusinersen. Bioanalysis. 2024. doi:10.4155/bio-2023-0218
BibTeX
@article{Zhang2024,
author = {Zhang, Xiao and Sha, Chunjie and Zhang, Wei and Zhao, Fengjuan and Zhu, Mingli and Leng, Guangyi and Liu, Wanhui},
title = {Development, Validation and Application of an Ion-Pair Reversed-Phase Liquid Chromatography-Tandem Mass Spectrometry Method for the Quantification of Nusinersen},
journal = {Bioanalysis},
year = {2024},
doi = {10.4155/bio-2023-0218}
}