Repeatability of Corneal Densitometry Measurements Using a Scheimpflug Camera in Healthy Normal Corneas

Mojgan Pakbin, Mehdi Khabazkhoob, Mohammad Pakravan, Akbar Fotouhi, Ebrahim Jafarzadehpur, Mohamadreza Aghamirsalim, Hassan Hashemi

Journal of Current Ophthalmology · 2022-01 · 34 (1) · 50-55

DOI: 10.4103/joco.joco_173_21 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Current Ophthalmology, 2022-01. Verfasst von Mojgan Pakbin und 6 weiteren Autor:innen. Beteiligte Einrichtung: Noor Ophthalmology Research Center, Noor Eye Hospital, Tehran, Iran. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To determine the repeatability of corneal densitometry measured by the Scheimpflug imaging system.

Methods

This cross-sectional study was conducted on photorefractive keratectomy candidates. One eye of each participant underwent imaging using Pentacam HR three times, 10 min apart. The repeatability of densitometry measurements was evaluated in four concentric annuli around the corneal apex and in different corneal depths. The repeatability of the measurements was evaluated using the intraclass correlation coefficient (ICC), repeatability coefficient (RC), and coefficient of variation (CV). The difference of repeatability between layers and zones was tested by tolerance index (TI).

Results

Sixty eyes of sixty patients with a mean age of 27.76 ± 3.93 years were studied. Half of the participants were female ( n = 30, 50%). ICC was above 0.9 in all corneal parts. The posterior layer and central zones showed the least variability of densitometry measurements considering the CV values. The RC was 2.06, 1.17, and 0.92 in anterior, central, and posterior layers, respectively. The RC was 0.88, 0.71, 1.51, and 4.56 in 0–2, 2–6, 6–10, and 10–12 mm circles, respectively. Only the reliability of densitometry in 10–12 mm annulus was statistically lower than the central zone (TI = 0.71).

Conclusions

Corneal densitometry measurements provided by the Pentacam had good repeatability. The repeatability of densitometry measurements decreased from the center to the periphery (with an exception for 0–2 mm and 2–6 mm) and from the posterior to the anterior of the cornea. The reliability of the 10–12 mm zone was markedly less than other zones.

Schlagwörter

Stichwörter der Autor:innen

Corneal densitometryReliabilityScheimpflug imaging system

Autor:innen und Institutionen

  • Noor Ophthalmology Research Center, Noor Eye Hospital, Tehran, Iran.
  • Translational Ophthalmology Research Center, Tehran University of Medical Sciences, Tehran, Iran.
  • Department of Basic Sciences, School of Nursing and Midwifery, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
  • Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
  • Department of Epidemiology and Biostatistics, School of Public Health, Tehran University of Medical Sciences, Tehran, Iran.
  • Noor Research Center for Ophthalmic Epidemiology, Noor Eye Hospital, Tehran, Iran.

Zitieren

APA

Pakbin, M., Khabazkhoob, M., Pakravan, M., Fotouhi, A., Jafarzadehpur, E., Aghamirsalim, M., & Hashemi, H. (2022). Repeatability of Corneal Densitometry Measurements Using a Scheimpflug Camera in Healthy Normal Corneas. Journal of Current Ophthalmology, 34(1), 50-55. https://doi.org/10.4103/joco.joco_173_21

Vancouver

Pakbin M, Khabazkhoob M, Pakravan M, Fotouhi A, Jafarzadehpur E, Aghamirsalim M, et al. Repeatability of Corneal Densitometry Measurements Using a Scheimpflug Camera in Healthy Normal Corneas. Journal of Current Ophthalmology. 2022;34(1):50-55. doi:10.4103/joco.joco_173_21

BibTeX
@article{Pakbin2022,
  author = {Pakbin, Mojgan and Khabazkhoob, Mehdi and Pakravan, Mohammad and Fotouhi, Akbar and Jafarzadehpur, Ebrahim and Aghamirsalim, Mohamadreza and Hashemi, Hassan},
  title = {Repeatability of Corneal Densitometry Measurements Using a Scheimpflug Camera in Healthy Normal Corneas},
  journal = {Journal of Current Ophthalmology},
  year = {2022},
  volume = {34},
  number = {1},
  pages = {50--55},
  doi = {10.4103/joco.joco_173_21}
}
Kostenlos starten →