Chemoport Catheter Fracture with an Unusual Site of Lodgement: A Case Report
Journal of Onco-Anaesthesiology and Perioperative Medicine · 2026-07 · 3 (Suppl 1) · S18-S21
DOI: 10.4103/joapm.joapm_90_25 ↗Kurz gefasst
Fallbericht in der Zeitschrift Journal of Onco-Anaesthesiology and Perioperative Medicine, online erschienen am 22.07.2026. Verfasst von Mallika Ganesh und 1 weiteren Autor:innen. Die Arbeit nennt 8 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Chemoport insertion is a widely accepted practice for administering chemotherapy as it provides a reliable long-term venous access. However, it is associated with several complications. These include early complications such as hematoma, air embolism and pneumothorax; and delayed complications such as infections, thrombosis, catheter fracture, migration and pinch-off syndrome. Catheter dislodgement is a rare but serious complication with an incidence of about 0.5%–1%. Here, we report a case where a part of the chemoport catheter got pinched off and migrated distally, which was incidentally noticed during chemoport removal. The fractured part was found at an unusual site, in the right great saphenous vein and was removed surgically. Insertion and handling of chemoports is a specialised procedure that needs expertise and experience. Once catheter fracture and dislodgement are recognised, urgent removal is recommended to prevent complications due to catheter embolisation.
Zitieren
APA
Ganesh, M., & Nallamad, V. K. (2026). Chemoport Catheter Fracture with an Unusual Site of Lodgement: A Case Report. Journal of Onco-Anaesthesiology and Perioperative Medicine, 3(Suppl 1), S18-S21. https://doi.org/10.4103/joapm.joapm_90_25
Vancouver
Ganesh M, Nallamad VK. Chemoport Catheter Fracture with an Unusual Site of Lodgement: A Case Report. Journal of Onco-Anaesthesiology and Perioperative Medicine. 2026;3(Suppl 1):S18-S21. doi:10.4103/joapm.joapm_90_25
BibTeX
@article{Ganesh2026,
author = {Ganesh, Mallika and Nallamad, Vinod Krishnamurthy},
title = {Chemoport Catheter Fracture with an Unusual Site of Lodgement: A Case Report},
journal = {Journal of Onco-Anaesthesiology and Perioperative Medicine},
year = {2026},
volume = {3},
number = {Suppl 1},
pages = {S18--S21},
doi = {10.4103/joapm.joapm_90_25}
}
EUVIMED