Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy
Journal of Medical Physics · 2023-07 · 48 (3) · 292-297
DOI: 10.4103/jmp.jmp_27_23 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Physics, online erschienen am 18.09.2023. Verfasst von Thamonwan Chirdchid und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Purpose
The focus of this work was given on the relative stopping power (RSP) using the water equivalent thickness (WET) validation on tissue substitutes and real pig organs, as well as a dosimetric comparison of proton treatment plans between single-energy computed tomography (SECT) and dual-energy computed tomography (DECT)-based dose calculations. Materials and Methods: The CT calibration curve of SECT and DECT data was generated using the stoichiometric calibration method. WET measurement was performed for RSP validation using a Giraffe dosimeter (IBA dosimetry) in various substitute tissues (Gammex) and real pig tissues. The thorax (008A, CIRS) and head (731-HN, CIRS) phantoms were used to generate proton plans. The dosimetric evaluations of SECT and DECT-based plans were performed using the gamma analysis with 1%/1 mm and the dose–volume histograms (DVHs) comparison.
Results
For RSP validation of substitute tissues, the largest percent WET difference between measurement and calculation was observed up to 17.9% (4 mm) in lung tissue, using SECT based. In real pig tissues, the average WET difference was 2.3% ± 2.1% and 2.5% ± 2.3% for SECT and DECT, respectively. The average gamma passed of about 92.1% for the lung and 96.8% for the head regions was reported. For the lung region, the DVH of the target dose was observed with a higher predicted dose in SECT than in DECT, while results in the head region were in good agreement for both SECT and DECT.
Conclusion
The performed dosimetric comparison indicates the dose differences between SECT and DECT. The impact of the CT calibration curve is more pronounced for the thorax region.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
- Her Royal Highness Princess Maha Chakri Sirindhorn Proton Center, Bangkok, Thailand.
- Department of Radiology, Division of Radiation Oncology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
- Department of Radiology, Division of Radiation Oncology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
Zitieren
APA
Chirdchid, T., Ruangchan, S., & Sanghangthum, T. (2023). Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy. Journal of Medical Physics, 48(3), 292-297. https://doi.org/10.4103/jmp.jmp_27_23
Vancouver
Chirdchid T, Ruangchan S, Sanghangthum T. Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy. Journal of Medical Physics. 2023;48(3):292-297. doi:10.4103/jmp.jmp_27_23
BibTeX
@article{Chirdchid2023,
author = {Chirdchid, Thamonwan and Ruangchan, Sirinya and Sanghangthum, Taweap},
title = {Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy},
journal = {Journal of Medical Physics},
year = {2023},
volume = {48},
number = {3},
pages = {292--297},
doi = {10.4103/jmp.jmp_27_23}
}
EUVIMED