Fachartikel in der Zeitschrift Journal of Medical Physics, online erschienen am 24.03.2025. Verfasst von S A Yoganathan und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, National Center for Cancer Care and Research, Hamad Medical Corporation, Doha, Qatar. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-SA 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Online adaptive radiation therapy (OART) poses unique challenges for quality assurance (QA), requiring innovative methodologies beyond traditional techniques. This study introduced an end-to-end (E2E) QA test for the Ethos OART system. Materials and Methods: Initial treatment plans were developed using deformed computed tomography (CT) images of standard phantoms. During treatment sessions, adaptive plans were created and delivered using undistorted physical QA phantoms equipped with measuring detectors. Our approach was demonstrated using standard QA phantoms – OCTAVIUS-four-dimensional (PTW, Freiburg, Germany), ArcCHECK (Sun Nuclear Corp., FL, USA), and the RUBY (PTW, Freiburg, Germany) – to evaluate the accuracy of contouring, synthetic CT (sCT), and dosimetry of adaptive plans in the Ethos OART system.

Results

Our findings demonstrated the superior performance of the Ethos OART system, with a gamma pass rate exceeding 96% (2% local/2 mm) and point dose deviations below 0.5%. The Dice coefficients for body contours between the sCT and reference CT were above 0.9, and the sCT accuracy was confirmed by mean absolute errors of

Schlagwörter

Stichwörter der Autor:innen

AdaptiveEthosend-to-endonlinequality assuranceradiotherapy

Autor:innen und Institutionen

  • Department of Radiation Oncology, National Center for Cancer Care and Research, Hamad Medical Corporation, Doha, Qatar.
  • Department of Radiation Oncology, Saint John Regional Hospital, Saint John, New Brunswick, Canada.

Zitieren

APA

Yoganathan, S. A., Khemissi, A., Paloor, S., Hammoud, R., & Al-Hammadi, N. (2025). An End-to-end Quality Assurance Procedure for Ethos Online Adaptive Radiotherapy. Journal of Medical Physics, 50(1), 140-147. https://doi.org/10.4103/jmp.jmp_188_24

Vancouver

Yoganathan SA, Khemissi A, Paloor S, Hammoud R, Al-Hammadi N. An End-to-end Quality Assurance Procedure for Ethos Online Adaptive Radiotherapy. Journal of Medical Physics. 2025;50(1):140-147. doi:10.4103/jmp.jmp_188_24

BibTeX
@article{Yoganathan2025,
  author = {Yoganathan, S A and Khemissi, Amine and Paloor, Satheesh and Hammoud, Rabih and Al-Hammadi, Noora},
  title = {An End-to-end Quality Assurance Procedure for Ethos Online Adaptive Radiotherapy},
  journal = {Journal of Medical Physics},
  year = {2025},
  volume = {50},
  number = {1},
  pages = {140--147},
  doi = {10.4103/jmp.jmp_188_24}
}
Kostenlos starten →