Biomechanical comparison of seven occipital–cervical fixation constructs using finite element modeling
Journal of Craniovertebral Junction and Spine · 2026-07 · 17 (4) · 384-391
DOI: 10.4103/jcvjs.jcvjs_85_26 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Craniovertebral Junction and Spine, online erschienen am 24.07.2026. Verfasst von Ian Polyzois und 6 weiteren Autor:innen. Beteiligte Einrichtung: Orthopaedic Innovation Centre, Winnipeg, MB, Canada. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Study Design: This was a basic science study.
Objective
The objective of this study was to determine which occipital–cervical (OC) fixation construct best balances construct stiffness with minimization of occipital screw pullout forces. Materials and Methods: Finite element analysis was used to simulate seven OC construct variations spanning across C1 with fixation in C2 and C3. Variations included the use of occiput rods, keel plates with varying width, two-hole clamps, two-hole clamps in addition to medial or lateral one-hole clamps, and occiput screw diameter. Modeling was in accordance with ASTM F2706/ISO 12189 guidelines. Materials were in accordance with ASTM F1295/ASTM F1839/ISO 10243 guidelines. Outcomes included bending stiffness, axial screw pullout reaction forces, and von Mises stress distribution to identify stress localization and predicted failure modes. Constructs were ranked using multiattribute utility theory (MAUT).
Results
Construct E (comprised two-hole clamps with added one-hole medial clamps) demonstrated the highest bending stiffness (820 N.mm/deg) and the highest MAUT utility score (0.79), representing the optimal balance of stiffness and minimized screw pullout forces. Construct F (comprised two-hole clamps alone) demonstrated the lowest bending stiffness (509 N.mm/deg).
Conclusions
A construct composed of medially fixed one-hole and two-hole clamps provided the best balance of bending stiffness and minimized occipital screw pullout reaction forces. Aligning screws near the median plane, using bilateral clamp fixation, and avoiding components that mechanically join the two rods improved construct performance. Increasing occipital screw diameter from 4.5 mm to 5.0 mm had little biomechanical impact. Stress localization at occipital component–rod interfaces suggests a potential weak point with risk of notching and fatigue failure.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Orthopaedic Innovation Centre, Winnipeg, MB, Canada.
- Max Rady College of Medicine, University of Manitoba, Winnipeg, MB, Canada.
- Division of Orthopaedic Surgery, University of Manitoba, Winnipeg, MB, Canada.
- Numalogics, Montreal, QC, Canada.
Zitieren
APA
Polyzois, I., Singh, J., Gustafson, S., Gascoyne, T., Le Navéaux, F., Legay, P., & Goytan, M. (2026). Biomechanical comparison of seven occipital–cervical fixation constructs using finite element modeling. Journal of Craniovertebral Junction and Spine, 17(4), 384-391. https://doi.org/10.4103/jcvjs.jcvjs_85_26
Vancouver
Polyzois I, Singh J, Gustafson S, Gascoyne T, Le Navéaux F, Legay P, et al. Biomechanical comparison of seven occipital–cervical fixation constructs using finite element modeling. Journal of Craniovertebral Junction and Spine. 2026;17(4):384-391. doi:10.4103/jcvjs.jcvjs_85_26
BibTeX
@article{Polyzois2026,
author = {Polyzois, Ian and Singh, Jaskaran and Gustafson, Sara and Gascoyne, Trevor and Le Navéaux, Franck and Legay, Pierre and Goytan, Michael},
title = {Biomechanical comparison of seven occipital–cervical fixation constructs using finite element modeling},
journal = {Journal of Craniovertebral Junction and Spine},
year = {2026},
volume = {17},
number = {4},
pages = {384--391},
doi = {10.4103/jcvjs.jcvjs_85_26}
}