Vergleichsstudie in der Zeitschrift Indian Journal of Ophthalmology, 2026. Verfasst von Gülsüm G Bozhöyük und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Republic of Turkey Ministry of Health Bitlis Tatvan State Hospital, Bitlis, Turkey. Hauptthemen laut MEDLINE: Visual Field Tests, Visual Fields, Virtual Reality, Pseudotumor Cerebri. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Purpose
To compare a newly developed virtual reality-based perimeter (Oculera visual field analyzer) with the Humphrey Field Analyzer (HFA) in patients with idiopathic intracranial hypertension (IIH).
Design
A cross-sectional study.
Methods
In this study, 30 eyes of 15 IIH patients were randomly tested with HFA (24-2, SITA Standard) and Oculera (24-2, Oculera Interactive) on the same visit. Test durations, global indices [visual field index (VFI), mean deviation (MD), and pattern standard deviation (PSD)], and false negative and positive user responses were compared. Spearman correlation was used for statistical analysis.
Results
Correlation analysis showed a statistically significant correlation between the measurements of the two devices in terms of the global indices (VFI, MD, and PSD) and test duration (r = 0.361, P = 0.050; r = 0.395, P = 0.031; r = 0.408, P = 0.025; and r = 0.363, P = 0.049, respectively). The Oculera was within an acceptable range for 83.3% of eyes, considering the MD reproducibility of 1.5–4 dB for HFA. No significant difference was found between the devices in false positive and negative responses ( P = 0.098 and P = 0.395, respectively).
Conclusions
Standard automated perimetry is the gold standard for detecting visual field defects, but it is costly and requires prolonged attention and stable fixation. The results of the Oculera, produced with virtual reality technology, were found to be consistent with those of the HFA, and its much lower cost and portability suggest that it has the potential to be an alternative to conventional perimeters.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Ophthalmology, Republic of Turkey Ministry of Health Bitlis Tatvan State Hospital, Bitlis, Turkey.
- Department of Ophthalmology, Ankara University Faculty of Medicine, Ankara, Turkey.
Zitieren
APA
Bozhöyük, G. G., Kızıltunç, P. B., & Atilla, H. (2026). Comparison of virtual reality-based portable perimetry and conventional perimetry in patients with idiopathic intracranial hypertension. Indian Journal of Ophthalmology. https://doi.org/10.4103/ijo.ijo_416_25
Vancouver
Bozhöyük GG, Kızıltunç PB, Atilla H. Comparison of virtual reality-based portable perimetry and conventional perimetry in patients with idiopathic intracranial hypertension. Indian Journal of Ophthalmology. 2026. doi:10.4103/ijo.ijo_416_25
BibTeX
@article{Bozhyk2026,
author = {Bozhöyük, Gülsüm G and Kızıltunç, Pınar B and Atilla, Huban},
title = {Comparison of virtual reality-based portable perimetry and conventional perimetry in patients with idiopathic intracranial hypertension},
journal = {Indian Journal of Ophthalmology},
year = {2026},
doi = {10.4103/ijo.ijo_416_25}
}
EUVIMED