EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access

Validation of a new technique using 4-0 nylon thread as a guide for tube insertion of the Ahmed glaucoma valve into the ciliary sulcus

Keisuke Nitta, Tomoko Miyakubo, Shunsuke Tokui, Mayuko Takayama, Yoichiro Shinohara, Kazuma Saito, Tomomi Endo, Hideo Akiyama

Indian Journal of Ophthalmology · 2024

DOI: 10.4103/ijo.ijo_3156_23 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Indian Journal of Ophthalmology, 2024. Verfasst von Keisuke Nitta und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Gunma University Graduate School of Medicine, 3-39-15 Showa-machi, Maebashi, Gunma, Japan. Hauptthemen laut MEDLINE: Glaucoma Drainage Implants, Ciliary Body, Glaucoma, Intraocular Pressure. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In this study, we validated the new technique for inserting the tube of the Ahmed glaucoma valve (AGV) into the ciliary sulcus. AGV tube insertion was performed by introducing a 4-0 nylon thread into the anterior chamber from the corneal incision on the opposite side of the AGV. The thread was placed in the lumen of a 23-G needle that was inserted into the sclera and ciliary sulcus from the AGV side. Withdrawing the 23-G needle, the thread was led out of the eye from the scleral wound, passing through the ciliary sulcus. The thread was then placed in the AGV tube lumen, and the tube was inserted through the scleral wound into the ciliary sulcus by using the thread as a guide, as per the Seldinger technique. In all seven eyes, the tube was precisely inserted into the ciliary sulcus without serious intraoperative complications such as vitreous straying of the tube.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Glaucoma Drainage ImplantsCiliary BodyGlaucomaIntraocular PressureNylonsHumansSuture TechniquesMaleFemaleProsthesis ImplantationMiddle AgedSuturesAged

Substanzen

Nylons

Autor:innen und Institutionen

  • Department of Ophthalmology, Gunma University Graduate School of Medicine, 3-39-15 Showa-machi, Maebashi, Gunma, Japan.
  • Medical Mycology Unit, Graduate School of Medicine, Teikyo University, 2-11-1 Kaga, Itabashi-ku, Tokyo, Japan.

Zitieren

APA

Nitta, K., Miyakubo, T., Tokui, S., Takayama, M., Shinohara, Y., Saito, K., Endo, T., & Akiyama, H. (2024). Validation of a new technique using 4-0 nylon thread as a guide for tube insertion of the Ahmed glaucoma valve into the ciliary sulcus. Indian Journal of Ophthalmology. https://doi.org/10.4103/ijo.ijo_3156_23

Vancouver

Nitta K, Miyakubo T, Tokui S, Takayama M, Shinohara Y, Saito K, et al. Validation of a new technique using 4-0 nylon thread as a guide for tube insertion of the Ahmed glaucoma valve into the ciliary sulcus. Indian Journal of Ophthalmology. 2024. doi:10.4103/ijo.ijo_3156_23

BibTeX
@article{Nitta2024,
  author = {Nitta, Keisuke and Miyakubo, Tomoko and Tokui, Shunsuke and Takayama, Mayuko and Shinohara, Yoichiro and Saito, Kazuma and Endo, Tomomi and Akiyama, Hideo},
  title = {Validation of a new technique using 4-0 nylon thread as a guide for tube insertion of the Ahmed glaucoma valve into the ciliary sulcus},
  journal = {Indian Journal of Ophthalmology},
  year = {2024},
  doi = {10.4103/ijo.ijo_3156_23}
}
Kostenlos starten →