Safety and therapeutic efficacy of 577 nm yellow solid krypton laser for retinopathy of prematurity; A prospective pilot study

Aneesha Vyas, Sarvesh Tiwari, Anand Subramanyam

Indian Journal of Ophthalmology · 2024-06 · 72 (6) · 856-859

DOI: 10.4103/ijo.ijo_2490_23 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Indian Journal of Ophthalmology, online erschienen am 05.02.2024. Verfasst von Aneesha Vyas und 2 weiteren Autor:innen. Beteiligte Einrichtung: Vitreo-Retina Department, K B H Bachooali Charitable Ophthalmic and ENT Hospital, Mumbai, Maharashtra, India. Hauptthemen laut MEDLINE: Retinopathy of Prematurity, Gestational Age, Laser Coagulation. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate the outcome of 577 nm yellow solid krypton laser as an independent treatment modality for Retinopathy of Prematurity.

Design

Prospective, non-comparative, non-randomized, interventional.

Methods

A prospective pilot study was conducted on 30 treatment eyes with retinopathy of prematurity (ROP) who underwent treatment with a 577 nm yellow solid state krypton laser. 3nethraNEO pediatric retinal imaging system was used to screen these infants born between March to August 2022. Pre procedure data such as gestational age (weeks), birth weight (g), and exposure to risk factors were recorded. Power (mW), exposure duration (msec), and number of laser spots administered were noted. Pain was assessed using the Neonatal Pain/Agitation and Sedation Scale (N-PASS). Post procedure data such as regression, persistence or reactivation of disease, and adverse effects were looked for. All eyes were followed up for a period of 3 months. Statistical Analysis Used: Chi-square test, Fisher’s exact test.

Results

Out of 30 eyes, 09 (30%) had type I and 21 (70%) had non-type I ROP. In type I disease, the mean power used was 88+/-10 mW with exposure duration of 130+/-27 msec and an average number of 890+/-260 burns were administered. In non-type I disease, the mean power used was 87+/-10 mW with an exposure duration of 150 ms and an average number of 900+/-290 burns were administered. The mean pain score calculated was 5 in type I and 6 in non-type I ROP. All eyes showed regression of disease after treatment.

Conclusion

577 nm yellow solid krypton laser was safe and effective in treating retinopathy of prematurity disease.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Retinopathy of PrematurityGestational AgeLaser CoagulationHumansPilot ProjectsProspective StudiesFemaleMaleInfant, NewbornTreatment OutcomeLasers, Solid-StateFollow-Up StudiesVisual Acuity

Autor:innen und Institutionen

  • Vitreo-Retina Department, K B H Bachooali Charitable Ophthalmic and ENT Hospital, Mumbai, Maharashtra, India.

Zitieren

APA

Vyas, A., Tiwari, S., & Subramanyam, A. (2024). Safety and therapeutic efficacy of 577 nm yellow solid krypton laser for retinopathy of prematurity; A prospective pilot study. Indian Journal of Ophthalmology, 72(6), 856-859. https://doi.org/10.4103/ijo.ijo_2490_23

Vancouver

Vyas A, Tiwari S, Subramanyam A. Safety and therapeutic efficacy of 577 nm yellow solid krypton laser for retinopathy of prematurity; A prospective pilot study. Indian Journal of Ophthalmology. 2024;72(6):856-859. doi:10.4103/ijo.ijo_2490_23

BibTeX
@article{Vyas2024,
  author = {Vyas, Aneesha and Tiwari, Sarvesh and Subramanyam, Anand},
  title = {Safety and therapeutic efficacy of 577 nm yellow solid krypton laser for retinopathy of prematurity; A prospective pilot study},
  journal = {Indian Journal of Ophthalmology},
  year = {2024},
  volume = {72},
  number = {6},
  pages = {856--859},
  doi = {10.4103/ijo.ijo_2490_23}
}