EUVIMEDEUVIMED entdecken
BeobachtungsstudieOpen Access

Performance of a handheld ultrasound device in detecting retinal detachments

Connor Nathe, Nicholas Max Scoville, Kenneth Matthew McKay, Kasra Rezaei, Shu Feng

Indian Journal of Ophthalmology · 2024

DOI: 10.4103/ijo.ijo_230_24 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Indian Journal of Ophthalmology, 2024. Verfasst von Connor Nathe und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, University of Washington School of Medicine, Seattle, WA, USA. Hauptthemen laut MEDLINE: Retinal Detachment, Ultrasonography. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 10 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To assess the accuracy and performance of the Butterfly IQ+ handheld ultrasound (HHUS) in detecting retinal detachments (RDs).

Methods

A cross-sectional observational study of eyes with (n = 20) and without (n = 20) RD imaged using a conventional B-scan ophthalmic ultrasound device Ellex Eye Cubed by Clarion Medical Technologies (Cambridge, ON, CA) and a portable HHUS (Butterfly IQ+). Images were compared between the modalities for qualitative differences. Agreement between the HHUS and standard ultrasonography equipment and with clinical examination was quantified using Cohen’s kappa coefficient.

Results

Forty eyes of 33 patients were included in the study. Twenty eyes of 18 patients had RDs, and 20 eyes of 18 patients did not. The observed agreement in RD diagnosis between the Butterfly and the Ellex Eye Cubed was 97.5%., and the Cohen’s Kappa was 0.950 (95% CI: 0.85–1.00). The observed agreement in RD diagnosis between the Butterfly IQ+ and diagnosis confirmed by the gold standard of clinical exam was 92.5%, and Cohen’s Kappa was 0.850 (95% CI: 0.69–1.00). The sensitivity and specificity of the Butterfly ultrasound in RD diagnosis were 90% (95% CI: 68.3–98.8%) and 95% (95% CI: 83.2–100), respectively.

Conclusion

There was a high degree of agreement between the systems for the identification of RD. The Butterfly IQ’s extreme portability and ease of use make it a viable option to screen for RDs in hospital-based consults, emergency departments, or low-resource settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Retinal DetachmentUltrasonographyHumansCross-Sectional StudiesMaleFemaleMiddle AgedReproducibility of ResultsAdultAgedEquipment DesignRetinaYoung Adult

Autor:innen und Institutionen

  • Department of Ophthalmology, University of Washington School of Medicine, Seattle, WA, USA.
  • Department of Ophthalmology, Ohio State College of Medicine, Columbus, OH, USA.
  • Department of Ophthalmology, Massachusetts Eye and Ear, Harvard Medical School, Boston, MA, USA.

Zitieren

APA

Nathe, C., Scoville, N. M., McKay, K. M., Rezaei, K., & Feng, S. (2024). Performance of a handheld ultrasound device in detecting retinal detachments. Indian Journal of Ophthalmology. https://doi.org/10.4103/ijo.ijo_230_24

Vancouver

Nathe C, Scoville NM, McKay KM, Rezaei K, Feng S. Performance of a handheld ultrasound device in detecting retinal detachments. Indian Journal of Ophthalmology. 2024. doi:10.4103/ijo.ijo_230_24

BibTeX
@article{Nathe2024,
  author = {Nathe, Connor and Scoville, Nicholas Max and McKay, Kenneth Matthew and Rezaei, Kasra and Feng, Shu},
  title = {Performance of a handheld ultrasound device in detecting retinal detachments},
  journal = {Indian Journal of Ophthalmology},
  year = {2024},
  doi = {10.4103/ijo.ijo_230_24}
}
Kostenlos starten →