Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
Indian Journal of Ophthalmology · 2023-05 · 71 (5) · 1927-1931
DOI: 10.4103/ijo.ijo_1982_22 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Indian Journal of Ophthalmology, online erschienen am 17.05.2023. Verfasst von Amit C Porwal und 3 weiteren Autor:innen. Beteiligte Einrichtung: Senior Consultant Cataract and Head Glaucoma Services, Department of Cataract and Glaucoma, Choithram Netralaya, Indore, Madhya Pradesh, India. Hauptthemen laut MEDLINE: Intraocular Pressure, Tonometry, Ocular. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-SA 4.0).
Abstract (Originalsprache)
Purpose
To compare readings of intraocular pressure (IOP) taken with the Goldmann applanation tonometer (GAT), the non-contact tonometer (NCT), and the rebound tonometer (RBT), and to compare their correlation with central corneal thickness (CCT).
Methods
This was a prospective, cross-sectional, observational study to which patients above 18 years of age were enrolled. A total of 400 eyes of 200 non-glaucomatous patients underwent IOP recordings using the GAT, NCT, and RBT, and CCT was also noted. Informed consent of the patients was taken. The IOP readings taken via the three methods were compared and correlated with CCT. Paired t test was used to compare the two devices. Simple and multivariate linear regression analyses were used to study the relationship between factors. A P value less than 0.05 was considered significant. Correlation was determined using the Pearson correlation coefficient, and a Bland–Altman graph was plotted.
Results
Mean IOP measured by the NCT was 15.65 ± 2.80 mmHg, by the RBT was 14.23 ± 3.05 mmHg, and by the GAT was 14.69 ± 2.97 mmHg. The mean CCT was 510.61 ± 33.83 microns. The difference between mean IOP recorded by the NCT and that by the RBT was 1.41 ± 2.39 mmHg, between the NCT and GAT was 0.95 ± 2.03 mmHg, and between the GAT and RBT was 0.45 ± 2.22 mmHg. The difference between the IO P values was statistically significant ( P < 0.005). All tonometers showed a statistically significant correlation with CCT, but it was observed that the NCT had a stronger correlation (0.4037).
Conclusion
The IOP readings taken by all the three methods were comparable; however, RBT values were closer to GAT values. CCT did influence the IOP values, and this should be kept in mind while evaluating.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Senior Consultant Cataract and Head Glaucoma Services, Department of Cataract and Glaucoma, Choithram Netralaya, Indore, Madhya Pradesh, India.
- Department of Community Ophthalmology, Choithram Netralaya, Indore, Madhya Pradesh, India.
- Department of Ophthalmology, Choithram Netralaya, Indore, Madhya Pradesh, India.
- Department of Cornea, Aravind Eye Hospital, Coimbatore, Tamil Nadu, India.
Zitieren
APA
Porwal, A. C., Shrishrimal, M., Punamia, R. P., & Mathew, B. C. (2023). Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness. Indian Journal of Ophthalmology, 71(5), 1927-1931. https://doi.org/10.4103/ijo.ijo_1982_22
Vancouver
Porwal AC, Shrishrimal M, Punamia RP, Mathew BC. Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness. Indian Journal of Ophthalmology. 2023;71(5):1927-1931. doi:10.4103/ijo.ijo_1982_22
BibTeX
@article{Porwal2023,
author = {Porwal, Amit C and Shrishrimal, Meghna and Punamia, Ravin P and Mathew, Bennet Chacko},
title = {Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness},
journal = {Indian Journal of Ophthalmology},
year = {2023},
volume = {71},
number = {5},
pages = {1927--1931},
doi = {10.4103/ijo.ijo_1982_22}
}
EUVIMED