Biodegradable material for glaucoma drainage devices – A pilot study in rabbits
Indian Journal of Ophthalmology · 2024
DOI: 10.4103/ijo.ijo_1092_23 ↗Kurz gefasst
Fachartikel in der Zeitschrift Indian Journal of Ophthalmology, 2024. Verfasst von Simar Rajan Singh und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Postgraduate Institute of Medical Education and Research, Chandigarh, India. Hauptthemen laut MEDLINE: Glaucoma Drainage Implants, Glaucoma, Absorbable Implants, Disease Models, Animal. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Purpose
To study the morphological and histological characteristics of the fibrous capsule formed around a novel biodegradable glaucoma drainage device (GDD) implant made of oxidized regenerated cellulose (ORC) after 1 and 3 months of its implantation.
Methods
It was a prospective, interventional, preclinical experimental study using New Zealand white rabbits. All animals underwent GDD implantation with a novel biodegradable device. The device’s body was made of ORC and attached to a silicone tube, connecting the body to the anterior chamber. Histopathology (hematoxylin and eosin/Masson’s trichrome staining) and immunohistochemistry (Alpha-Smooth Muscle Actin expression) characteristics of the bleb formed around the novel device were noted at 1 and 3 months.
Results
One month post implantation, the biodegradable material produced a significant foreign body type of reaction evidenced by the exuberant infiltration by macrophages, lymphocytes, and multinucleated giant cells. The granulomatous response subsided by 3 months with disorganized collagen deposition on Masson’s trichrome staining. The silicone tube was an internal control, and histopathology demonstrated well-organized collagen deposition around it at 3 months. Immunohistochemistry for α-smooth muscle actin also demonstrated more myofibroblast transformation at the site of the tube than the biodegradable implant.
Conclusions
Our results indicate that the tissue response around a biodegradable GDD was different from the response to conventional devices. A diffuse, loose vascularized mesh was observed to develop, which may be more compatible with sustained IOP control over a longer period, in contrast to the usual thick and dense capsule formed around the non-biodegradable devices.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Ophthalmology, Postgraduate Institute of Medical Education and Research, Chandigarh, India.
- Department of Ophthalmology, University of Iowa Hospitals and Clinics, Iowa City, Iowa, USA.
Zitieren
APA
Singh, S. R., Pandav, S. S., Kaushik, S., Nada, R., Gautam, N., Kaur, S., & Thattaruthody, F. (2024). Biodegradable material for glaucoma drainage devices – A pilot study in rabbits. Indian Journal of Ophthalmology. https://doi.org/10.4103/ijo.ijo_1092_23
Vancouver
Singh SR, Pandav SS, Kaushik S, Nada R, Gautam N, Kaur S, et al. Biodegradable material for glaucoma drainage devices – A pilot study in rabbits. Indian Journal of Ophthalmology. 2024. doi:10.4103/ijo.ijo_1092_23
BibTeX
@article{Singh2024,
author = {Singh, Simar Rajan and Pandav, Surinder Singh and Kaushik, Sushmita and Nada, Ritambhra and Gautam, Natasha and Kaur, Savleen and Thattaruthody, Faisal},
title = {Biodegradable material for glaucoma drainage devices – A pilot study in rabbits},
journal = {Indian Journal of Ophthalmology},
year = {2024},
doi = {10.4103/ijo.ijo_1092_23}
}
EUVIMED