Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation

Farhad Ahamed, Debkumar Pal, Sarika Palepu, Jeevanmuktha Somashekara, Sibasish Sahoo, Ayan Roy, Tanmay Saha, Semanti Das, Kalyan Goswami

Indian Journal of Community Medicine · 2026-06 · 51 (Suppl 3) · S517-S523

DOI: 10.4103/ijcm.ijcm_932_25 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Indian Journal of Community Medicine, online erschienen am 18.07.2026. Verfasst von Farhad Ahamed und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Community Medicine and Family Medicine, All India Institute of Medical Sciences, Kalyani, West Bengal, India. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

We aimed to conduct this study to compare the validity of the HemoCue® Glucose 201 RT System (point-of-care testing (POCT) device) in estimating blood glucose levels in patients attending a tertiary care hospital, considering the automated analyzer method (Vitros® 4600 chemistry system, QuidelOrtho Corporation, USA) as the gold standard.

Methods

We conducted this study among the patients attending the outpatient department in a tertiary care hospital in Eastern India. Venous blood was collected to estimate fasting blood glucose (FBS) and postprandial blood glucose (PPBS) using the automated analyzer and the POCT device, along with capillary blood to estimate FBS and PPBS using the POCT device.

Results

We enrolled 200 patients with a mean age of 49.4 (standard deviation 12.6) years. Lin’s CCC for POCT device compared to automated analyzer (venous blood) for FBS estimation was 0.95 (95% confidence interval (CI): 0.94 to 0.96, P value < 0.001) for venous blood and 0.95 (95% CI: 0.94–0.97, P value < 0.001) for capillary blood. Lin’s CCC for PPBS estimation was 0.99 (95% CI: 0.98–0.99, P value < 0.001) for venous blood and 0.98 (95% CI: 0.97–0.98, P value < 0.001) for capillary blood.

Conclusion

The POCT device showed substantial concordance with the gold standard method for measuring both FBS and PPBS. It also demonstrated a fair agreement with the automated analyzer in estimating blood glucose levels.

Schlagwörter

Stichwörter der Autor:innen

Diagnostic accuracyblood glucosepoint-of-care testingvalidation

Autor:innen und Institutionen

  • Department of Community Medicine and Family Medicine, All India Institute of Medical Sciences, Kalyani, West Bengal, India.
  • Department of Biochemistry, All India Institute of Medical Sciences, Kalyani, West Bengal, India.
  • Department of Endocrinology and Metabolism, All India Institute of Medical Sciences, Kalyani, West Bengal, India.
  • Centre for Community Medicine, All India Institute of Medical Sciences, New Delhi, India.

Zitieren

APA

Ahamed, F., Pal, D., Palepu, S., Somashekara, J., Sahoo, S., Roy, A., Saha, T., Das, S., & Goswami, K. (2026). Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation. Indian Journal of Community Medicine, 51(Suppl 3), S517-S523. https://doi.org/10.4103/ijcm.ijcm_932_25

Vancouver

Ahamed F, Pal D, Palepu S, Somashekara J, Sahoo S, Roy A, et al. Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation. Indian Journal of Community Medicine. 2026;51(Suppl 3):S517-S523. doi:10.4103/ijcm.ijcm_932_25

BibTeX
@article{Ahamed2026,
  author = {Ahamed, Farhad and Pal, Debkumar and Palepu, Sarika and Somashekara, Jeevanmuktha and Sahoo, Sibasish and Roy, Ayan and Saha, Tanmay and Das, Semanti and Goswami, Kalyan},
  title = {Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation},
  journal = {Indian Journal of Community Medicine},
  year = {2026},
  volume = {51},
  number = {Suppl 3},
  pages = {S517--S523},
  doi = {10.4103/ijcm.ijcm_932_25}
}
Kostenlos starten →