Portable vs. Fixed X-Ray Devices: An Analytical Cross-Sectional Study on the Dentist’s Thyroid Gland Radiation Exposure
Dental Hypotheses · 2026-10 · 17 (4) · 50-52
DOI: 10.4103/denthyp.denthyp_68_25 ↗Kurz gefasst
Querschnittstudie in der Zeitschrift Dental Hypotheses, online erschienen am 02.11.2026. Verfasst von Ali Hakiem Tawfieq und 1 weiteren Autor:innen. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Introduction
Occupational exposure from preapical X‑rays may elevate thyroid radiation dose in dentists. This study compared thyroid gland exposure resulting from portable vs. fixed intra‑oral X‑ray devices.
Methods
A cross‑sectional field study measured thyroid dose by thermoluminescent dosimeters placed on the anterior neck (thyroid level) using portable and fixed oral X‑ray units during routine preapical imaging.
Results
We found a statistically significant difference regarding mean thyroid dose (mGy) between the fixed devices group (n=30, mean = 0.00219 ± 3.84e − 4) and portable devices group ( n = 30, mean = 0.00536 ± 9.58e − 4) ( P < 0.001). Linear regression model (overall model test: P < 0.001) showed the device type and radiographic technique were significant predictors for thyroid dose ( P < 0.05), while collimation technique and usage of thyroid collar were not significant predictors ( P > 0.05).
Conclusions
Portable units pose a higher thyroid exposure risk to dentists.
Zitieren
APA
Tawfieq, A. H., & Hasan, H. A. (2026). Portable vs. Fixed X-Ray Devices: An Analytical Cross-Sectional Study on the Dentist’s Thyroid Gland Radiation Exposure. Dental Hypotheses, 17(4), 50-52. https://doi.org/10.4103/denthyp.denthyp_68_25
Vancouver
Tawfieq AH, Hasan HA. Portable vs. Fixed X-Ray Devices: An Analytical Cross-Sectional Study on the Dentist’s Thyroid Gland Radiation Exposure. Dental Hypotheses. 2026;17(4):50-52. doi:10.4103/denthyp.denthyp_68_25
BibTeX
@article{Tawfieq2026,
author = {Tawfieq, Ali Hakiem and Hasan, Haider Ali},
title = {Portable vs. Fixed X-Ray Devices: An Analytical Cross-Sectional Study on the Dentist’s Thyroid Gland Radiation Exposure},
journal = {Dental Hypotheses},
year = {2026},
volume = {17},
number = {4},
pages = {50--52},
doi = {10.4103/denthyp.denthyp_68_25}
}