When Impedance Trend Lies, Remote Monitoring Helps for Diagnosis
Annals of Clinical Cardiology · 2023
DOI: 10.4103/accj.accj_10_23 ↗Kurz gefasst
Fachartikel in der Zeitschrift Annals of Clinical Cardiology, 2023. Verfasst von Marco Scicchitano und 9 weiteren Autor:innen. Die Arbeit nennt 9 Literaturangaben.
Abstract (Originalsprache)
We described the fracture of an ICD lead in a patient at the third device replacement.Through the remote monitoring electric interference was detected on the defibrillator lead despite the pacing and shock impedances during the leadlifetime remained stable and in a normal range. X-ray confirmed the lead failure and patient underwent to a lead replacement.
Zitieren
APA
Scicchitano, M., Schirripa, V., Loricchio, M. L., Castro, A., Iulianella, R. V., Perna, S., Centurion, C. A., Albano, B., Rizzo, V., & Granatelli, A. (2023). When Impedance Trend Lies, Remote Monitoring Helps for Diagnosis. Annals of Clinical Cardiology. https://doi.org/10.4103/accj.accj_10_23
Vancouver
Scicchitano M, Schirripa V, Loricchio ML, Castro A, Iulianella RV, Perna S, et al. When Impedance Trend Lies, Remote Monitoring Helps for Diagnosis. Annals of Clinical Cardiology. 2023. doi:10.4103/accj.accj_10_23
BibTeX
@article{Scicchitano2023,
author = {Scicchitano, Marco and Schirripa, Valentina and Loricchio, Maria Luisa and Castro, Antonio and Iulianella, Renzo Venanzio and Perna, Silvia and Centurion, Carlos Aznaran and Albano, Bruno and Rizzo, Veronica and Granatelli, Antonino},
title = {When Impedance Trend Lies, Remote Monitoring Helps for Diagnosis},
journal = {Annals of Clinical Cardiology},
year = {2023},
doi = {10.4103/accj.accj_10_23}
}