The relationship between compression force, image quality and radiation dose in mammography
South African Journal of Radiology · 30.11.2009 · 13 (4) · 86
DOI: 10.4102/sajr.v13i4.486 ↗Kurz gefasst
Fachartikel in der Zeitschrift South African Journal of Radiology, online erschienen am 30.11.2009. Verfasst von A Korf und 2 weiteren Autor:innen. Die Arbeit wurde bisher 9-mal zitiert (Crossref). Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Mammography aims to obtain mammograms of best possible image quality with least possible radiation dose.1 Theoretically, an increase in breast compression gives a reduction in thickness, without changing the density, resulting in improved image quality and reduced radiation dose.
Aim
This study aims to investigate the relationship between compression force, phantom thickness, image quality and radiation dose. The existence of a compression point beyond which increased compression gives a change in density rather than thickness is also considered.
Method
Image quality is assessed with a contrast-detail phantom within Superflab phantom on a computed radiography (CR) mammography unit using automatic exposure control (AEC). Image quality is determined by visual inspection and image quality figure (IQF) scoring. The effect of compression and lesion depth on image quality is determined. Entrance and exit doses are calculated. The relationship between entrance dose, compression and thickness is investigated, as is the existence of a compression point beyond which a change in phantom density occurs. The average glandular dose (AGD) is calculated from the scanning average level (SAL) and logarithmic mean (LgM) according to Koen et al,2 and compared to the allowable limit.
Results
The geometry effect was not observed. An improvement in image quality with increased compression was found. Entrance dose did decrease with increased compression. This trend was not observed with exit dose as AEC was used and exit dose was calculated from SAL values. The “change-in-density” point of compression was determined. Both LgM and SAL could be used successfully for AGD calculation.
Zitieren
APA
Korf, A., Herbst, C. P., & Rae, W. I. (2009). The relationship between compression force, image quality and radiation dose in mammography. South African Journal of Radiology, 13(4), 86. https://doi.org/10.4102/sajr.v13i4.486
Vancouver
Korf A, Herbst CP, Rae WI. The relationship between compression force, image quality and radiation dose in mammography. South African Journal of Radiology. 2009;13(4):86. doi:10.4102/sajr.v13i4.486
BibTeX
@article{Korf2009,
author = {Korf, A and Herbst, C P and Rae, W ID},
title = {The relationship between compression force, image quality and radiation dose in mammography},
journal = {South African Journal of Radiology},
year = {2009},
volume = {13},
number = {4},
pages = {86},
doi = {10.4102/sajr.v13i4.486}
}
EUVIMED