Fachartikel in der Zeitschrift Monaldi Archives for Chest Disease, online erschienen am 30.07.2025. Verfasst von Abdias Rodriguez und 6 weiteren Autor:innen. Beteiligte Einrichtung: Harvard Medical School, Boston, MA. Hauptthemen laut MEDLINE: Thoracentesis, Drainage, Pleural Effusion. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Equipoise exists regarding the optimal method to drain pleural fluid during thoracentesis. While several institutions use wall-based automated suction, others point to the risk of excessively high suction pressures and therefore elevated barotrauma risk as a reason to avoid it. We first performed in vitro experiments involving drainage of a 1-liter saline bag using standard thoracentesis apparatus, a digital manometer, and either manual drainage (using a 60 mL syringe) or automated drainage (using wall suction at the maximum setting). The proceduralist was blinded to measurements during manual aspiration. Separately, in a clinical setting involving consecutive hospitalized adults undergoing thoracentesis, dynamic suction pressures were similarly measured during automated drainage. Total aspirated volume, time-to-evacuation, patient discomfort, and complications were also recorded. In vitro experiments showed that compared to manual aspiration, automated drainage using wall suction resulted in shorter average time-to-evacuation (230 sec vs. 365 sec), lower suction pressures (average maximum: -361±4.5 cmH2O vs. -496±5.1 cmH2O, p
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Harvard Medical School, Boston, MA.
- Division of Pulmonary and Critical Care Medicine, Brigham and Women's Hospital, Harvard Medical School, Boston, MA.
- Harvard Medical School, Boston, MA; Division of Pulmonary and Critical Care Medicine, Brigham and Women's Hospital, Harvard Medical School, Boston, MA.
Zitieren
APA
Rodriguez, A., Ramaswamy, A., Menon, A. A., Belkhayat, N., Forth, V. E., Schissel, S. L., & Shafiq, M. (2025). Suction pressures generated during thoracentesis using wall suction-based automated drainage: an in vitro and in vivo analysis. Monaldi Archives for Chest Disease, 96(4). https://doi.org/10.4081/monaldi.2025.3484
Vancouver
Rodriguez A, Ramaswamy A, Menon AA, Belkhayat N, Forth VE, Schissel SL, et al. Suction pressures generated during thoracentesis using wall suction-based automated drainage: an in vitro and in vivo analysis. Monaldi Archives for Chest Disease. 2025;96(4). doi:10.4081/monaldi.2025.3484
BibTeX
@article{Rodriguez2025,
author = {Rodriguez, Abdias and Ramaswamy, Anuradha and Menon, Aravind A and Belkhayat, Noah and Forth, Victoria E and Schissel, Scott L and Shafiq, Majid},
title = {Suction pressures generated during thoracentesis using wall suction-based automated drainage: an in vitro and in vivo analysis},
journal = {Monaldi Archives for Chest Disease},
year = {2025},
volume = {96},
number = {4},
doi = {10.4081/monaldi.2025.3484}
}
EUVIMED