EUVIMEDEUVIMED entdecken
Crossref

Reliability analysis of GEM® Premier™ technology: a multicenter study

Francesco Faggiano, Tiziana Franchin, Matteo Ritrovato, Pietro Derrico

Emergency Care Journal · 2015

DOI: 10.4081/ecj.2015.4795 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Emergency Care Journal, 2015. Verfasst von Francesco Faggiano und 3 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper describes how specific blood gas analyzer characteristics can support the emerging clinical and organizational user’s needs while ensuring patient safety. A oneyear data set from two Italian hospitals was analyzed from 10 different blood gas analyzers. Performance measurements in terms of mean down time (MDT) were calculated to show how technical characteristics declared by the manufacturer compare to the analyzer availability in real clinical settings. Results show a high level of reliability for the analyzed technology, associated with very low MDT of each sensor integrated in the cartridge. Moreover, results show a high level of efficiency in cartridge usage. Such results are consistent with the specification of the GEM® Premier™ maintenance- free technology and are particularly relevant in the point-of-care testing setting.

Zitieren

APA

Faggiano, F., Franchin, T., Ritrovato, M., & Derrico, P. (2015). Reliability analysis of GEM® Premier™ technology: a multicenter study. Emergency Care Journal. https://doi.org/10.4081/ecj.2015.4795

Vancouver

Faggiano F, Franchin T, Ritrovato M, Derrico P. Reliability analysis of GEM® Premier™ technology: a multicenter study. Emergency Care Journal. 2015. doi:10.4081/ecj.2015.4795

BibTeX
@article{Faggiano2015,
  author = {Faggiano, Francesco and Franchin, Tiziana and Ritrovato, Matteo and Derrico, Pietro},
  title = {Reliability analysis of GEM® Premier™ technology: a multicenter study},
  journal = {Emergency Care Journal},
  year = {2015},
  doi = {10.4081/ecj.2015.4795}
}
Kostenlos starten →