EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY-NC 4.0

Enhanced Rotational Stability with the Locking Anchor EndoButton System: A Biomechanical Comparison of Free Bone Graft Non-Screw Methods for Anterior Shoulder Instability

Muhammed Furkan Tosun iD, Ethem Burak Oklaz iD, İbrahim Kaya iD, Fatma Kübra Erbay Elibol iD, Elif Naz Perdeci iD, Teyfik Demir iD, Tolga Tolunay iD, Ulunay Kanatlı iD

Clinics in Orthopedic Surgery · 2025

DOI: 10.4055/cios24459 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Clinics in Orthopedic Surgery, 2025. Verfasst von Muhammed Furkan Tosun und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics and Traumatology, Yıldırım Beyazıt University Yenimahalle Training and Research Hospital, Ankara, Türkiye. Hauptthemen laut MEDLINE: Joint Instability, Bone Transplantation, Shoulder Joint, Suture Anchors. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Backgroud: Despite the increasing use of non-screw free bone graft fixation methods for glenoid defects associated with anterior shoulder instability, biomechanical knowledge of their effectiveness remains limited. This study aims to evaluate the biomechanical stability of different non-screw iliac crest graft fixation methods for anterior shoulder instability.

Methods

Forty-eight scapula bone models were used to compare the mechanical properties of non-screw fixation methods. The amples were divided into 4 groups: group 1 with 2 anchors, group 2 with 2 EndoButtons, group 3 with a combination of 2 anchors and 1 EndoButton, and group 4 with the Locking Anchor EndoButton System (LAES). Following fixation, dynamic testing began with a 10-N preload. After the preload, cyclic loading was applied between 10 N and 150 N for 100 cycles. Static loading was then conducted at a rate of 0.08 mm/sec. The test continued until either system failure occurred or the graft displacement reached 7 mm. Stiffness and maximum load values were determined. Rotational tests were performed at an angular velocity of 10°/sec. Torsional stiffness and maximum torque capacity were determined.

Results

Under dynamic loading conditions, the LAES group exhibited significantly less displacement compared to all other groups, while no significant differences were observed among the remaining groups. In static testing, the LAES group demonstrated the highest stiffness (183.354 ± 33.295 N/mm), significantly exceeding that of the Anchor group (93.847 ± 20.401 N/mm) and Anchor-EndoButton group (135.036 ± 37.815 N/mm). The EndoButton and LAES groups withstood significantly higher maximum loads than the Anchor and Anchor-EndoButton groups, with the LAES group achieving the highest load capacity (543.333 ± 100.851 N). In rotational testing, the LAES group displayed significantly greater torsional stiffness (0.182 ± 0.063 N·m/°) compared to all other groups. Additionally, the LAES group withstood the highest maximum torque (3.068 ± 1.194 N·m), significantly surpassing the Anchor, EndoButton, and Anchor-EndoButton groups.

Conclusions

All tested fixation methods provided sufficient stability under early postoperative loading conditions. However, the LAES demonstrated superior biomechanical performance, particularly in terms of rotational stability, compared to other non-screw fixation techniques.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Joint InstabilityBone TransplantationShoulder JointSuture AnchorsHumansBiomechanical PhenomenaRotationScapula

Stichwörter der Autor:innen

BiomechanicsFree bone graftGlenohumeral instabilityGlenoid reconstruction

Autor:innen und Institutionen

  • Department of Orthopaedics and Traumatology, Yıldırım Beyazıt University Yenimahalle Training and Research Hospital, Ankara, Türkiye.
  • Department of Orthopaedics and Traumatology, Gazi University Faculty of Medicine, Ankara, Türkiye.
  • Department of Orthopaedics and Traumatology, Dr. Abdurrahman Yurtaslan Ankara Oncology Training and Research Hospital, Ankara, Türkiye.
  • Department of Biomedical Engineering, TOBB University of Economics and Technology, Ankara, Türkiye.

ORCID-Profile: Muhammed Furkan Tosun, Ethem Burak Oklaz, İbrahim Kaya, Fatma Kübra Erbay Elibol, Elif Naz Perdeci, Teyfik Demir, Tolga Tolunay, Ulunay Kanatlı

Zitieren

APA

Tosun, M. F., Oklaz, E. B., Kaya, İ., Erbay Elibol, F. K., Perdeci, E. N., Demir, T., Tolunay, T., & Kanatlı, U. (2025). Enhanced Rotational Stability with the Locking Anchor EndoButton System: A Biomechanical Comparison of Free Bone Graft Non-Screw Methods for Anterior Shoulder Instability. Clinics in Orthopedic Surgery. https://doi.org/10.4055/cios24459

Vancouver

Tosun MF, Oklaz EB, Kaya İ, Erbay Elibol FK, Perdeci EN, Demir T, et al. Enhanced Rotational Stability with the Locking Anchor EndoButton System: A Biomechanical Comparison of Free Bone Graft Non-Screw Methods for Anterior Shoulder Instability. Clinics in Orthopedic Surgery. 2025. doi:10.4055/cios24459

BibTeX
@article{Tosun2025,
  author = {Tosun, Muhammed Furkan and Oklaz, Ethem Burak and Kaya, İbrahim and Erbay Elibol, Fatma Kübra and Perdeci, Elif Naz and Demir, Teyfik and Tolunay, Tolga and Kanatlı, Ulunay},
  title = {Enhanced Rotational Stability with the Locking Anchor EndoButton System: A Biomechanical Comparison of Free Bone Graft Non-Screw Methods for Anterior Shoulder Instability},
  journal = {Clinics in Orthopedic Surgery},
  year = {2025},
  doi = {10.4055/cios24459}
}
Kostenlos starten →