A comparative biomechanical study of original and compatible titanium bases: evaluation of screw loosening and 3D-crown displacement following cyclic loading analysis

Rimantas Ožiūnas iD, Jurgina Sakalauskienė iD, Darius Jegelevičius iD, Gintaras Janužis iD

The Journal of Advanced Prosthodontics · 2022 · 14 (2) · 70

DOI: 10.4047/jap.2022.14.2.70 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Advanced Prosthodontics, 2022. Verfasst von Rimantas Ožiūnas und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics, Medical Academy, Lithuanian University of Health Sciences, Kaunas, Lithuania. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: This study evaluated screw loosening and 3D crown displacement after cyclic loading of implant-supported incisor crowns cemented with original titanium bases or with three compatible, nonoriginal components. MATERIALS AND METHODS: A total of 32 dental implants were divided into four groups (n = 8 each): Group 1 used original titanium bases, while Groups 2-4 used compatible components. The reverse torque value (RTV) was evaluated prior to and after cyclic loading (1,200,000 cycles). Samples (prior to and after cyclic loading) were scanned with a microcomputed tomography (micro-CT). Preload and postload files were superimposed by 3D inspection software, and 3D crown displacement analysis was performed using root-mean-square (RMS) values. All datasets were analyzed using one-way ANOVA and Tukey's post hoc analysis. RESULTS: Significant variations were observed in the postload RTV, depending on the titanium base brand (P < .001). The mean postload RTVs were significantly higher in Groups 1 and 2 than in the other study groups. While evaluating 3D crown displacement, the lowest mean RMS value was shown in the original Group 1, with the highest RMS value occurring in Group 4. CONCLUSION: Within the limitations of this in vitro study and under the implemented conditions, it was concluded that the manufacturer brand of the titanium base significantly influenced screw loosening following the fatigue test and influenced 3D crown displacement after cyclic loading.

Schlagwörter

Stichwörter der Autor:innen

CrownDental abutmentDental implantTorqueX-ray microtomography

Autor:innen und Institutionen

  • Department of Prosthodontics, Medical Academy, Lithuanian University of Health Sciences, Kaunas, Lithuania.
  • Faculty of Electrical and Electronics Engineering, Biomedical Engineering Institute, Kaunas University of Technology, Kaunas, Lithuania.
  • Department of Maxillofacial Surgery, Lithuanian University of Health Sciences, Kaunas, Lithuania.

ORCID-Profile: Rimantas Ožiūnas, Jurgina Sakalauskienė, Darius Jegelevičius, Gintaras Janužis

Zitieren

APA

Ožiūnas, R., Sakalauskienė, J., Jegelevičius, D., & Janužis, G. (2022). A comparative biomechanical study of original and compatible titanium bases: evaluation of screw loosening and 3D-crown displacement following cyclic loading analysis. The Journal of Advanced Prosthodontics, 14(2), 70. https://doi.org/10.4047/jap.2022.14.2.70

Vancouver

Ožiūnas R, Sakalauskienė J, Jegelevičius D, Janužis G. A comparative biomechanical study of original and compatible titanium bases: evaluation of screw loosening and 3D-crown displacement following cyclic loading analysis. The Journal of Advanced Prosthodontics. 2022;14(2):70. doi:10.4047/jap.2022.14.2.70

BibTeX
@article{Oinas2022,
  author = {Ožiūnas, Rimantas and Sakalauskienė, Jurgina and Jegelevičius, Darius and Janužis, Gintaras},
  title = {A comparative biomechanical study of original and compatible titanium bases: evaluation of screw loosening and 3D-crown displacement following cyclic loading analysis},
  journal = {The Journal of Advanced Prosthodontics},
  year = {2022},
  volume = {14},
  number = {2},
  pages = {70},
  doi = {10.4047/jap.2022.14.2.70}
}
Kostenlos starten →