Identification of factors affecting the retention of 99mTc-DMSA in syringes among the paediatric population in Sri Lanka
Sri Lanka Journal of Medical Sciences · 29.03.2026 · 1 (2)
DOI: 10.4038/sljms1.v1i2.14 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sri Lanka Journal of Medical Sciences, online erschienen am 29.03.2026. Verfasst von Vishakha Paramee und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
99mTc-DMSA is a common paediatric renal scintigraphy study, where radiopharmaceutical is introduced intravenously using disposable syringes. Radiopharmaceuticals can be retained on disposable syringes, causing under-dosing, and impacting imaging quality in paediatric patients. The study aimed to specify the factors affecting on retention of 99mTc-DMSA in disposable syringes. Furthermore, the influence of the medical staff, patient positioning, and the ascendancy of the MON.TEK 99mMo/99mTc generator were investigated.
Methods
A study of 212 pediatric patients undergoing 99mTc-DMSA renal scintigraphy was conducted using CAPINTEC-CRC-15R and CAPINTEC-CRC/15 BETA radiation dose calibrators. The 99mTc-Pertechnetate was eluted with a MON.TEK 99mMo/99mTc generator and the radiopharmaceutical volume were recorded. Six types of 1ml/1cc disposable syringes (Types 1-6) and one 1ml insulin syringe (Type 7) were used. Pre- and post-doses of these syringes were measured.
Results
Results showed a significant moderate, positive correlation between retention dose and dilution (radiopharmaceutical volume). There was no correlation between retention dose and pre-dose or patient weight. No significant difference in retention was found based on patient gender or position. Significant differences in retention dose were observed among syringe types, injection sites, labeling personnel, and administering personnel.
Conclusion
The retention of 99mTc-DMSA varies by syringe type, with Syringe Type-1 showing the highest retention and Type-7 the lowest, making Type-7 the most suitable. Retention is influenced by factors such as radiopharmaceutical dilution, labeling, injecting personnel, and injection site, while pre-dose, patient gender, weight, positioning, and decay time do not significantly affect retention.
Zitieren
APA
Paramee, V., & Wimalarathne, D. N. (2026). Identification of factors affecting the retention of 99mTc-DMSA in syringes among the paediatric population in Sri Lanka. Sri Lanka Journal of Medical Sciences, 1(2). https://doi.org/10.4038/sljms1.v1i2.14
Vancouver
Paramee V, Wimalarathne DN. Identification of factors affecting the retention of 99mTc-DMSA in syringes among the paediatric population in Sri Lanka. Sri Lanka Journal of Medical Sciences. 2026;1(2). doi:10.4038/sljms1.v1i2.14
BibTeX
@article{Paramee2026,
author = {Paramee, Vishakha and Wimalarathne, Dilani Neranjana},
title = {Identification of factors affecting the retention of 99mTc-DMSA in syringes among the paediatric population in Sri Lanka},
journal = {Sri Lanka Journal of Medical Sciences},
year = {2026},
volume = {1},
number = {2},
doi = {10.4038/sljms1.v1i2.14}
}
EUVIMED