Point-of-Care Potassium Measurement vs Artificial Intelligence–Enabled Electrocardiography for Hyperkalemia Detection
American Journal of Critical Care · 01.01.2025 · 34 (1) · 41-51
DOI: 10.4037/ajcc2025597 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift American Journal of Critical Care, erschienen am 01.01.2025. Verfasst von Chin Lin und 6 weiteren Autor:innen. Beteiligte Einrichtung: Chin Lin is an associate professor, School of Medicine, National Defense Medical Center, Taipei, Taiwan, Republic of China. Hauptthemen laut MEDLINE: Hyperkalemia, Electrocardiography, Artificial Intelligence, Potassium. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 39 Literaturangaben.
Abstract (Originalsprache)
Background
Hyperkalemia can be detected by point-of-care (POC) blood testing and by artificial intelligence– enabled electrocardiography (ECG). These 2 methods of detecting hyperkalemia have not been compared.
Objective
To determine the accuracy of POC and ECG potassium measurements for hyperkalemia detection in patients with critical illness.
Methods
This retrospective study involved intensive care patients in an academic medical center from October 2020 to September 2021. Patients who had 12-lead ECG, POC potassium measurement, and central laboratory potassium measurement within 1 hour were included. The POC potassium measurements were obtained from arterial blood gas analysis; ECG potassium measurements were calculated by a previously developed deep learning model. Hyperkalemia was defined as a central laboratory potassium measurement of 5.5 mEq/L or greater.
Results
Fifteen patients with hyperkalemia and 252 patients without hyperkalemia were included. The POC and ECG potassium measurements were available about 35 minutes earlier than central laboratory results. Correlation with central laboratory potassium measurement was better for POC testing than for ECG (mean absolute errors of 0.211 mEq/L and 0.684 mEq/L, respectively). For POC potassium measurement, area under the receiver operating characteristic curve (AUC) to detect hyperkalemia was 0.933, sensitivity was 73.3%, and specificity was 98.4%. For ECG potassium measurement, AUC was 0.884, sensitivity was 93.3%, and specificity was 63.5%.
Conclusions
The ECG potassium measurement, with its high sensitivity and coverage rate, may be used initially and followed by POC potassium measurement for rapid detection of life-threatening hyperkalemia.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Chin Lin is an associate professor, School of Medicine, National Defense Medical Center, Taipei, Taiwan, Republic of China.
- Chien-Chou Chen is a resident physician, Division of Nephrology, Department of Medicine, Tri-Service General Hospital, National Defense Medical Center, Taipei.
- Chin-Sheng Lin is a professor, Division of Cardiology, Department of Internal Medicine, Tri-Service General Hospital, National Defense Medical Center, Taipei.
- Hung-Sheng Shang is an associate professor, Division of Clinical Pathology, Department of Pathology, Tri-Service General Hospital, National Defense Medical Center, Taipei.
- Chia-Cheng Lee is an assistant professor, Department of Medical Informatics, and visiting staff, Division of Colorectal Surgery, Department of Surgery, Tri-Service General Hospital, National Defense Medical Center, Taip…
- Tom Chau is a medical director, Department of Medicine, Providence St Vincent Medical Center, Portland, Oregon.
Zitieren
APA
Lin, C., Chen, C. C., Lin, C. S., Shang, H. S., Lee, C. C., Chau, T., & Lin, S. H. (2025). Point-of-Care Potassium Measurement vs Artificial Intelligence–Enabled Electrocardiography for Hyperkalemia Detection. American Journal of Critical Care, 34(1), 41-51. https://doi.org/10.4037/ajcc2025597
Vancouver
Lin C, Chen CC, Lin CS, Shang HS, Lee CC, Chau T, et al. Point-of-Care Potassium Measurement vs Artificial Intelligence–Enabled Electrocardiography for Hyperkalemia Detection. American Journal of Critical Care. 2025;34(1):41-51. doi:10.4037/ajcc2025597
BibTeX
@article{Lin2025,
author = {Lin, Chin and Chen, Chien-Chou and Lin, Chin-Sheng and Shang, Hung-Sheng and Lee, Chia-Cheng and Chau, Tom and Lin, Shih-Hua},
title = {Point-of-Care Potassium Measurement vs Artificial Intelligence–Enabled Electrocardiography for Hyperkalemia Detection},
journal = {American Journal of Critical Care},
year = {2025},
volume = {34},
number = {1},
pages = {41--51},
doi = {10.4037/ajcc2025597}
}
EUVIMED