EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Quantitative assessment of bone density: correlation between Hounsfield units in multislice CT and gray values in cone-beam CT in selected maxillofacial regions

Haider Ali Hasan, Ali Hakiem Tawfieq, Rabeia J Khalil, Hayder Mahdi Idan

Romanian Journal of Stomatology · 30.06.2026 · 72 (2) · 155

DOI: 10.37897/rjs.2026.2.4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Romanian Journal of Stomatology, online erschienen am 11.06.2026. Verfasst von Haider Ali Hasan und 3 weiteren Autor:innen. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Accurate assessment of bone density is important for diagnosis, treatment planning, and implant placement in maxillofacial practice. Multislice computed tomography (MSCT) provides standardized bone density measurements using Hounsfield units (HU), whereas cone-beam computed tomography (CBCT) offers lower radiation exposure but provides non-standardized gray values (GV).

Objective

To assess the relationship between Hounsfield units obtained from MSCT and gray values obtained from CBCT across selected maxillofacial regions.

Methods

A total of 60 patients underwent MSCT and CBCT imaging for clinically indicated maxillofacial evaluation. Bone density measurements were performed at selected anatomical sites using standardized regions of interest. Statistical analysis included Pearson correlation, paired t-tests, Bland-Altman analysis, and linear regression.

Results

Strong positive correlations were observed between MSCT HU and CBCT GV across the analyzed regions (r = 0.85-0.91, p < 0.001). However, Bland-Altman analysis demonstrated wide limits of agreement (−145.2 to +108.0), indicating that the two modalities are not directly interchangeable.

Conclusion

CBCT gray values may reflect relative bone density trends but cannot be considered a direct quantitative substitute for MSCT Hounsfield units without appropriate calibration. CBCT may be useful for relative bone density assessment in maxillofacial practice, provided its limitations are recognized.

Zitieren

APA

Hasan, H. A., Tawfieq, A. H., Khalil, R. J., & Idan, H. M. (2026). Quantitative assessment of bone density: correlation between Hounsfield units in multislice CT and gray values in cone-beam CT in selected maxillofacial regions. Romanian Journal of Stomatology, 72(2), 155. https://doi.org/10.37897/rjs.2026.2.4

Vancouver

Hasan HA, Tawfieq AH, Khalil RJ, Idan HM. Quantitative assessment of bone density: correlation between Hounsfield units in multislice CT and gray values in cone-beam CT in selected maxillofacial regions. Romanian Journal of Stomatology. 2026;72(2):155. doi:10.37897/rjs.2026.2.4

BibTeX
@article{Hasan2026,
  author = {Hasan, Haider Ali and Tawfieq, Ali Hakiem and Khalil, Rabeia J and Idan, Hayder Mahdi},
  title = {Quantitative assessment of bone density: correlation between Hounsfield units in multislice CT and gray values in cone-beam CT in selected maxillofacial regions},
  journal = {Romanian Journal of Stomatology},
  year = {2026},
  volume = {72},
  number = {2},
  pages = {155},
  doi = {10.37897/rjs.2026.2.4}
}
Kostenlos starten →