Temperature Recording In Neonates by Non-Contact Thermometer Versus Routine Digital Thermometer - The Best Modality?
International Journal of Nursing Education · 22.10.2025 · 17 (4) · 42-49
DOI: 10.37506/pprzmf60 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Nursing Education, online erschienen am 22.10.2025. Verfasst von Paulami Chakraborty und 4 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref). Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Introduction
Measurement of temperature in neonates should be accurate, reliable and reproducible, considering that they have a limited ability to regulate their own temperature. There is no ideal thermometer to measure temperature, however, there has been a constant endeavour to find more accurate and non-invasive means of measurement. The present study was conducted to find whether Non-Contact Infrared thermometers were as accurate as Digital Axillary thermometers in measuring temperatures in neonates. Methodology: This was a comparative diagnostic study, carried out in the maternity unit of a tertiary care hospital over a period of 6 weeks. A total of 154 neonates were recruited in the study through consecutive sampling and their temperatures recorded using Non-Contact Infrared Thermometer (NCIT) on forehead and Digital Axillary Thermometer (DAT) on axilla simultaneously.
Results
The average temperature recording on forehead using NCIT and axilla using DAT, was in the range of 36 – 37oC, with the NCIT showing a higher recording as compared to the DAT. The mean difference between the two modalities was -0.46oC. A Bland-Altman plot analysis was done, in which majority of readings had wide variation, falling beyond the upper (0.57oC) and lower (-1.49oC) limits. A linear regression plot also showed a positive correlation between NCIT on forehead and DAT on axilla, with absolute R value = 0.2, which showed a weak correlation between the two.
Conclusion
It was concluded that NCIT measurements on forehead are not as accurate as DAT measurements on axilla. However, a larger study, is needed to extrapolate these findings to the general population.
Conclusion
It was concluded that NCIT measurements on forehead are not as accurate as DAT measurements on axilla. However, a larger study, is needed to extrapolate these findings to the general population.
Zitieren
APA
Chakraborty, P., AS, R., R, A. K., Shah, A., & Swain, S. (2025). Temperature Recording In Neonates by Non-Contact Thermometer Versus Routine Digital Thermometer - The Best Modality?. International Journal of Nursing Education, 17(4), 42-49. https://doi.org/10.37506/pprzmf60
Vancouver
Chakraborty P, AS R, R AK, Shah A, Swain S. Temperature Recording In Neonates by Non-Contact Thermometer Versus Routine Digital Thermometer - The Best Modality?. International Journal of Nursing Education. 2025;17(4):42-49. doi:10.37506/pprzmf60
BibTeX
@article{Chakraborty2025,
author = {Chakraborty, Paulami and AS, Reena and R, Aruna K and Shah, Ankur and Swain, Santosh},
title = {Temperature Recording In Neonates by Non-Contact Thermometer Versus Routine Digital Thermometer - The Best Modality?},
journal = {International Journal of Nursing Education},
year = {2025},
volume = {17},
number = {4},
pages = {42--49},
doi = {10.37506/pprzmf60}
}
EUVIMED