Evaluation of a digital fully-contactless optical device to quantify and record breathing pattern components at rest and after exercise

Panagiotis Sakkatos

Exploration of Digital Health Technologies · 2025

DOI: 10.37349/edht.2025.101156 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Exploration of Digital Health Technologies, 2025. Verfasst von Panagiotis Sakkatos. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aim

Structured light plethysmography (SLP), is a contactless optical system developed to monitor breathing patterns by analyzing chest-wall movement. It has not been thoroughly validated against other non-invasive motion analysis systems under different breathing conditions. This study therefore aimed to evaluate the criterion-validity of the SLP compared to the respiratory inductive plethysmography (RIP) at rest and after exercise.

Methods

Adults underwent two simultaneous 5-minute recordings from both devices, conducted at rest and following submaximal exercise on a cycle ergometer. Timing indices and thoracoabdominal (TA) movement parameters were examined. Measurement agreement between SLP and RIP was assessed using Bland-Altman plots at rest, after exercise, and for exercise-induced changes.

Results

Fifty adults (mean age 29.3 ± 6.8 years; 30 males) participated. Α total of 3,395 and 4,295 breath cycles were analyzed at rest and post-exercise, respectively. Over 92% of differences in timing parameters under both conditions were within the 95% limits of agreement (LOA) and their mean differences were found close to zero across a wide range of breath cycle magnitudes (rest: 2.62–8.06 s; post-exercise: 2.16–6.16 s). For ΤΑ movement parameters, the mean bias between devices at rest was 0.31 for ribcage amplitude (RCampi) and 0.23 for abdominal amplitude (ABampi), with LOA ranging from −0.06 to 0.66 and −0.06 to 0.52, respectively. A trend towards greater discrepancies for the individual measurements of RCampi and ABampi at higher magnitudes of TA movements was noted, especially post-exercise. A good average agreement between the devices was found for RCampi/ABampi both at rest [mean difference: 0.03, standard deviation (SD): 0.21] and after exercise (mean difference: 1.10, SD: 0.24).

Conclusions

The SLP is an accurate method to quantify and measure timing indices and the ratio of the ribcage motion to the abdominal motion under different breathing conditions.

Zitieren

APA

Sakkatos, P. (2025). Evaluation of a digital fully-contactless optical device to quantify and record breathing pattern components at rest and after exercise. Exploration of Digital Health Technologies. https://doi.org/10.37349/edht.2025.101156

Vancouver

Sakkatos P. Evaluation of a digital fully-contactless optical device to quantify and record breathing pattern components at rest and after exercise. Exploration of Digital Health Technologies. 2025. doi:10.37349/edht.2025.101156

BibTeX
@article{Sakkatos2025,
  author = {Sakkatos, Panagiotis},
  title = {Evaluation of a digital fully-contactless optical device to quantify and record breathing pattern components at rest and after exercise},
  journal = {Exploration of Digital Health Technologies},
  year = {2025},
  doi = {10.37349/edht.2025.101156}
}
Kostenlos starten →