EUVIMEDEUVIMED entdecken
FallberichtOpen Access · CC BY 4.0

Inadvertent implantable cardioverter defibrillator lead malpositioning in the left ventricle via a patent foramen ovale: a case report and review of the literature

Samrina Jannat, Ahmed Umer, Azeem S Sheikh

Exploration of Cardiology · 23.04.2026 · 4

DOI: 10.37349/ec.2026.1012104 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Exploration of Cardiology, online erschienen am 23.04.2026. Verfasst von Samrina Jannat und 2 weiteren Autor:innen. Die Arbeit nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Implantable cardioverter defibrillators (ICDs) play a central role in the prevention of sudden cardiac death. Although implantation is generally safe, rare but clinically significant complications may occur. One such complication is inadvertent placement of a transvenous ICD lead into the left ventricle via a patent foramen ovale (PFO), which carries a risk of systemic thromboembolism and may remain clinically silent for prolonged periods. We report a case of an ICD lead malpositioned in the left ventricle through a PFO, detected 17 months after implantation during transthoracic echocardiography performed for worsening dyspnoea. Imaging demonstrated the lead traversing the interatrial septum and mitral valve into the left ventricle. The patient underwent successful transvenous extraction and reimplantation of a new right ventricular lead without complications. This case highlights the limitations of electrocardiographic clues alone and underscores the importance of multimodality imaging and meticulous implantation technique. A focused narrative review integrates current literature regarding mechanisms, diagnostic strategies, management options, and preventive considerations.

Zitieren

APA

Jannat, S., Umer, A., & Sheikh, A. S. (2026). Inadvertent implantable cardioverter defibrillator lead malpositioning in the left ventricle via a patent foramen ovale: a case report and review of the literature. Exploration of Cardiology, 4. https://doi.org/10.37349/ec.2026.1012104

Vancouver

Jannat S, Umer A, Sheikh AS. Inadvertent implantable cardioverter defibrillator lead malpositioning in the left ventricle via a patent foramen ovale: a case report and review of the literature. Exploration of Cardiology. 2026;4. doi:10.37349/ec.2026.1012104

BibTeX
@article{Jannat2026,
  author = {Jannat, Samrina and Umer, Ahmed and Sheikh, Azeem S},
  title = {Inadvertent implantable cardioverter defibrillator lead malpositioning in the left ventricle via a patent foramen ovale: a case report and review of the literature},
  journal = {Exploration of Cardiology},
  year = {2026},
  volume = {4},
  doi = {10.37349/ec.2026.1012104}
}