Design and Analysis of Rehabilitation Training Mechanism Based on Upper Limb Rehabilitation Movements

Haoyu Han, Qinghui Zhu, Li Feng

Acta of Bioengineering and Biomechanics · 23.04.2026 · 28 (numer ciągły...)

DOI: 10.37190/abb/220943 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta of Bioengineering and Biomechanics, online erschienen am 23.04.2026. Verfasst von Haoyu Han und 2 weiteren Autor:innen. Beteiligte Einrichtung: College of Home and Art Design, Northeast Forestry University, Harbin, Heilongjiang, 150000, China. Hauptthemen laut MEDLINE: Upper Extremity.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To conduct a comparative analysis of the trajectory and joint torque differences between exoskeleton-type and end-effector-type upper limb rehabilitation training mechanisms during rehabilitation movements, and to identify the more suitable mechanism type for passive upper limb rehabilitation training.

Methods

Based on the Brunnstrom rehabilitation approach, upper limb rehabilitation movements for the passive training phase were selected. Muscle force simulations were conducted using OpenSim software to identify movements with high muscular engagement. The structures of the two mechanisms were designed through degree-of-freedom analysis, and motion trajectory simulations were completed in ADAMS. In MATLAB, discrete Fréchet distance is used to evaluate trajectory similarity. Calculate the error of comparing the trajectories of two mechanisms with the trajectory of the human body model at the same time using R2. In ADAMS, compare the rotational torque changes of the shoulder, elbow, and wrist joints between two mechanisms.

Results

The movement trajectory of the exoskeleton mechanism closely matched the natural motion of the human upper limb, error ≤ 2 mm, trajectory R2 ≥ 0.9975. Its joint torque distribution was balanced, with a maximum torque ≤ 40 Nm. The trajectory error of the end-effector mechanism ≤5 mm, trajectory R2 ≥ 0.9910 and torques were concentrated at the wrist and elbow joints, some values ≥ 40 Nm.

Conclusions

The exoskeleton-type upper limb rehabilitation training mechanism demonstrates superior performance in terms of movement trajectory adaptability and joint protection compared to the end-effector-type mechanism, making it more suitable for passive movement assistance and joint range of motion recovery training.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Upper ExtremityHumansTorqueMovementBiomechanical PhenomenaExoskeleton DeviceRange of Motion, ArticularComputer SimulationEquipment DesignModels, Biological

Stichwörter der Autor:innen

Mechanism analysisMechanism designRehabilitation deviceRehabilitation trainingTrajectory fitting

Autor:innen und Institutionen

  • College of Home and Art Design, Northeast Forestry University, Harbin, Heilongjiang, 150000, China.

Zitieren

APA

Han, H., Zhu, Q., & Feng, L. (2026). Design and Analysis of Rehabilitation Training Mechanism Based on Upper Limb Rehabilitation Movements. Acta of Bioengineering and Biomechanics, 28(numer ciągły...). https://doi.org/10.37190/abb/220943

Vancouver

Han H, Zhu Q, Feng L. Design and Analysis of Rehabilitation Training Mechanism Based on Upper Limb Rehabilitation Movements. Acta of Bioengineering and Biomechanics. 2026;28(numer ciągły...). doi:10.37190/abb/220943

BibTeX
@article{Han2026,
  author = {Han, Haoyu and Zhu, Qinghui and Feng, Li},
  title = {Design and Analysis of Rehabilitation Training Mechanism Based on Upper Limb Rehabilitation Movements},
  journal = {Acta of Bioengineering and Biomechanics},
  year = {2026},
  volume = {28},
  number = {numer ciągły...},
  doi = {10.37190/abb/220943}
}
Kostenlos starten →