Shear rate induced viscosity change of human blood samples and blood mimicking fluids

Eszter Szabó, Ernő Zsolt Baka, Kornél Tamás

Acta of Bioengineering and Biomechanics · 2024

DOI: 10.37190/abb-02405-2024-03 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta of Bioengineering and Biomechanics, 2024. Verfasst von Eszter Szabó und 2 weiteren Autor:innen. Beteiligte Einrichtung: 1Department of Machine and Product Design, Faculty of Mechanical Engineering, Budapest University of Technology and Economics, Budapest, Hungary. Hauptthemen laut MEDLINE: Blood Viscosity. Die Arbeit wurde bisher 3-mal zitiert (Crossref).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The aim of this study was to measure blood and blood mimicking fluids viscosity at different shear rates (on the interval of 0.1-5000 1/s and 0.1-10000 1/s) while taking into consideration the measuring device’s capability and blood’s characteristics. We also provided the measurement results of the most accurate measuring program.

Methods

We measured blood samples from five donors, and four different blood mimicking fluid compositions. The measurements were done on an Anton Paar Physica MCR301 rotational rheometer with two measuring programs varying in the shear rate intervals, the number of measuring points and the measuring point durations.

Results

The results confirmed the significant shear thinning and thixotropic effects of blood. Blood mimicking fluids also had these characteristics. The measured blood viscosity values are in agreement with those of the literature.

Conclusions

It can be concluded that the step test program was able to give more stable results as the measured torque was over the nominal limit of 0.05 μNm over 0.1 1/s and over the selected torque limit of 0.5 μNm over 31.6 1/s. Blood mimicking fluid measurement results were different from that of the literature due to different measuring conditions. The sample consisting of water, glycerol and starch mimicked well blood’s behaviour and viscosity values at 37 degrees Celsius.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood ViscosityHumansShear StrengthStress, MechanicalViscosity

Stichwörter der Autor:innen

bloodblood mimicking fluidsmeasurementsrheologythixotropyviscosity

Autor:innen und Institutionen

  • 1Department of Machine and Product Design, Faculty of Mechanical Engineering, Budapest University of Technology and Economics, Budapest, Hungary.

Zitieren

APA

Szabó, E., Baka, E. Z., & Tamás, K. (2024). Shear rate induced viscosity change of human blood samples and blood mimicking fluids. Acta of Bioengineering and Biomechanics. https://doi.org/10.37190/abb-02405-2024-03

Vancouver

Szabó E, Baka EZ, Tamás K. Shear rate induced viscosity change of human blood samples and blood mimicking fluids. Acta of Bioengineering and Biomechanics. 2024. doi:10.37190/abb-02405-2024-03

BibTeX
@article{Szab2024,
  author = {Szabó, Eszter and Baka, Ernő Zsolt and Tamás, Kornél},
  title = {Shear rate induced viscosity change of human blood samples and blood mimicking fluids},
  journal = {Acta of Bioengineering and Biomechanics},
  year = {2024},
  doi = {10.37190/abb-02405-2024-03}
}
Kostenlos starten →