EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Longitudinal slit in the distal portion of the abdominal catheter: a technique to minimize obstructions or malfunction in ventriculoperitoneal shunts

Marcelo Moraes Valença, Marcelo Andrade Valença, Martina Falcão Valença, Alex Caetano de Barros

Neurological Surgery and Anatomy · 2024

DOI: 10.37085/nsa.2024.11 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Neurological Surgery and Anatomy, 2024. Verfasst von Marcelo Moraes Valença und 3 weiteren Autor:innen. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

IntroductionOne of the primary reasons for ventriculoperitoneal (VP) shunt failure related to the abdominal catheter is the formation of a pseudocyst due to intraperitoneal septations. Often, it is not merely a blockage but an encasement of the catheter's tip within a closed cavity that restricts the absorption of cerebrospinal fluid (CSF), leading to pseudocyst formation. Such accumulation increases pressure, potentially exceeding the valve's threshold and causing valve malfunction, which can result in the recurrence of hydrocephalus and neurological symptoms.ObjectiveThis paper revisits the historical development of a technique pioneered by neurosurgeon Alex Caetano de Barros, which aims to reduce failures in ventriculoperitoneal shunts.TechniqueThe procedure involves making one or two longitudinal incisions of approximately 3 cm each on the abdominal catheter before insertion into the peritoneal cavity, positioned 15 cm and 10 cm from the catheter's distal end. Incisions are made on only one side of the catheter. Surgeons must be cautious when handling the catheter to avoid accidents with the scalpel.ConclusionThis simple technique can enhance the performance of VP shunts by potentially preventing complications due to obstruction at the distal end of the catheter, thereby facilitating better CSF flow and reducing the risk of shunt failure.

Zitieren

APA

Valença, M. M., Valença, M. A., Valença, M. F., & Barros, A. C. D. (2024). Longitudinal slit in the distal portion of the abdominal catheter: a technique to minimize obstructions or malfunction in ventriculoperitoneal shunts. Neurological Surgery and Anatomy. https://doi.org/10.37085/nsa.2024.11

Vancouver

Valença MM, Valença MA, Valença MF, Barros ACD. Longitudinal slit in the distal portion of the abdominal catheter: a technique to minimize obstructions or malfunction in ventriculoperitoneal shunts. Neurological Surgery and Anatomy. 2024. doi:10.37085/nsa.2024.11

BibTeX
@article{Valena2024,
  author = {Valença, Marcelo Moraes and Valença, Marcelo Andrade and Valença, Martina Falcão and Barros, Alex Caetano de},
  title = {Longitudinal slit in the distal portion of the abdominal catheter: a technique to minimize obstructions or malfunction in ventriculoperitoneal shunts},
  journal = {Neurological Surgery and Anatomy},
  year = {2024},
  doi = {10.37085/nsa.2024.11}
}
Kostenlos starten →