AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To assess agreement between corneal epithelial thickness maps obtained with MS-39 and Optovue Avanti OCT devices in healthy eyes, and to ascertain if their measures are interchangeable. Study Design: Descriptive comparative study. Place and Duration of Study: Amanat Eye Hospital, Rawalpindi from May 2025 to July 2025.

Methods

In this study, 123-eyes were scanned with MS-39 and Optovue. Central (0-2mm), paracentral (2-5mm) and mid-peripheral (5-7mm) epithelial thickness measurements were taken. The Shapiro-Wilk test was used to evaluate the normality of data. The pair measurements were compared using Wilcoxon signed-rank test. Reliability was determined with intraclass correlation coefficient (ICC) and agreement with Bland-Altman analysis. A p

Zitieren

APA

Bibi, A., Asrar, U., Akhtar, M., Ali, W., & Zahid, A. (2026). Comparison of Corneal Epithelial Thickness Mapping using Two Different Optical Coherence Tomography. Pakistan Journal of Ophthalmology, 42(3). https://doi.org/10.36351/pjo.v42i3.2241

Vancouver

Bibi A, Asrar U, Akhtar M, Ali W, Zahid A. Comparison of Corneal Epithelial Thickness Mapping using Two Different Optical Coherence Tomography. Pakistan Journal of Ophthalmology. 2026;42(3). doi:10.36351/pjo.v42i3.2241

BibTeX
@article{Bibi2026,
  author = {Bibi, Ayesha and Asrar, Umair and Akhtar, Maria and Ali, Waqas and Zahid, Arooj},
  title = {Comparison of Corneal Epithelial Thickness Mapping using Two Different Optical Coherence Tomography},
  journal = {Pakistan Journal of Ophthalmology},
  year = {2026},
  volume = {42},
  number = {3},
  doi = {10.36351/pjo.v42i3.2241}
}
Kostenlos starten →