The design of radiology viewing box using light emitting diode and potentiometer
GHMJ (Global Health Management Journal) · 2017
DOI: 10.35898/ghmj-1196 ↗Kurz gefasst
Fachartikel in der Zeitschrift GHMJ (Global Health Management Journal), 2017. Verfasst von Anak Agung Aris Diartama und 9 weiteren Autor:innen.
Abstract (Originalsprache)
Background
In the process of work to gain the maximum results, a radiologist needs a viewing box tool to read radiographs. Therefore, the authors want to develop a viewing box tool, which in general the work if this tool resembles the factory manufactured tool. The viewing tool box made can adjust the intensity of the light produced.Objective: to create a viewing box tool by using a potentiometer system.
Methods
This study used applied research method by creating and using the design of viewing box tool by using a potentiometer system and testing the viewing box tool created by using a Lux meter and 15 respondents consisting of five radiologists and 10 radiographers who should fulfill the questionnaire form.
Results
The mean of viewing box illumination reached 220 lux. The results of the questionnaire showed that 100% radiologist gave an A (excellent) and expressed that the viewing box tool created could be used properly and 90% radiographers provided an A (excellent) and expressed that the viewing box tool created could be used properly, while 10% radiographer gave a value of B (moderate).Conclusion: viewing box tool created could be used properly and obtained optimal results as a tool in reading radiographs. Potentiometer system contained in the viewing box was very helpful in reading radiographs because it allowed to adjust the light intensity according to user needs.Keywords : Viewing box, Potentiometer Bibliography : 1980-2011
Zitieren
APA
Diartama, A. A. A., Suswaty, S., Priantoro, W., Sugiyanto, S., Sudiyono, S., Anwar, M. C., Latifah, L., Santjaka, A., Amri, F., & Mulyantoro, D. K. (2017). The design of radiology viewing box using light emitting diode and potentiometer. GHMJ (Global Health Management Journal). https://doi.org/10.35898/ghmj-1196
Vancouver
Diartama AAA, Suswaty S, Priantoro W, Sugiyanto S, Sudiyono S, Anwar MC, et al. The design of radiology viewing box using light emitting diode and potentiometer. GHMJ (Global Health Management Journal). 2017. doi:10.35898/ghmj-1196
BibTeX
@article{Diartama2017,
author = {Diartama, Anak Agung Aris and Suswaty, Susy and Priantoro, Win and Sugiyanto, Sugiyanto and Sudiyono, Sudiyono and Anwar, M. Choiroel and Latifah, Leny and Santjaka, Aris and Amri, Faisal and Mulyantoro, Donny Kristanto},
title = {The design of radiology viewing box using light emitting diode and potentiometer},
journal = {GHMJ (Global Health Management Journal)},
year = {2017},
doi = {10.35898/ghmj-1196}
}
EUVIMED