Design of Reflectance Pulse Oximeter and BPM using the Max30100 Sensor in Early Detection of Hypoxemia in Patients with Cardiovascular Disorders
International Journal of Advanced Health Science and Technology · 2021
DOI: 10.35882/ijahst.v1i1.1 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Advanced Health Science and Technology, 2021. Verfasst von Chiu-Hua Huang und 1 weiteren Autor:innen. Die Arbeit wurde bisher 7-mal zitiert (Crossref). Der Volltext ist frei zugänglich (Open Access, CC BY-SA 4.0).
Abstract (Originalsprache)
Blood oxygen saturation meter is a tool used to monitor the state of oxygen saturation in the blood and also the patient's heart rate (BPM) and to assist in the physical assessment of the patient without going through blood gas analysis. Oxygen saturation measuring devices usually use the difference in the wavelengths of red and infrared led light that will be captured by the photodiode. The purpose of this research is to make a pulse oximeter equipped with a display of SPO2, BPM values and an additional SP02 signal. The design of this measuring instrument uses the MAX30100 sensor, the minimum system circuit of Arduino ATmega328p and OLED (Organic Light-Emitting Diode). Data from the MAX30100 sensor enters the I2C pin on the minimum Arduino system, then the microcontroller is processed to produce the percentage of SPO2 value, BPM value, and SPO2 signal which is then displayed on the OLED. The test is done by comparing the module with standard measuring instruments which produces the largest % error of 0.81% for Spo2 and 0.87% for BPM. The error presentation is obtained from factor measurements, if there is finger movement it will cause a large error. From the results obtained, the tool is still feasible to use because in the "Guidelines for Testing and Calibrating Medical Devices" Ministry of Health RI 2001, the maximum limit in the pulse oximeter error tolerance is for Spo2 1% and BPM 5%.
Zitieren
APA
Huang, C., & Guo, J. (2021). Design of Reflectance Pulse Oximeter and BPM using the Max30100 Sensor in Early Detection of Hypoxemia in Patients with Cardiovascular Disorders. International Journal of Advanced Health Science and Technology. https://doi.org/10.35882/ijahst.v1i1.1
Vancouver
Huang C, Guo J. Design of Reflectance Pulse Oximeter and BPM using the Max30100 Sensor in Early Detection of Hypoxemia in Patients with Cardiovascular Disorders. International Journal of Advanced Health Science and Technology. 2021. doi:10.35882/ijahst.v1i1.1
BibTeX
@article{Huang2021,
author = {Huang, Chiu-Hua and Guo, Jia-Wei},
title = {Design of Reflectance Pulse Oximeter and BPM using the Max30100 Sensor in Early Detection of Hypoxemia in Patients with Cardiovascular Disorders},
journal = {International Journal of Advanced Health Science and Technology},
year = {2021},
doi = {10.35882/ijahst.v1i1.1}
}
EUVIMED