EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

2,3‐Pentanedione – Method for the determination of 2,3‐pentanedione in workplace air using high performance liquid chromatography (HPLC)

Schuh, Claudia, Nitschke, Lutz, Hebisch, Ralph, Brock, Thomas Helmut, Hartwig, Andrea, MAK Commission

The MAK Collection for Occupational Health and Safety · 2025 · 3 (3)

DOI: 10.34865/am60014e1918a_w ↗

Kurz gefasst

Fachartikel in der Zeitschrift The MAK Collection for Occupational Health and Safety, 2025. Verfasst von Schuh und 10 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This analytical method is a validated measurement procedure for the determination of 2,3‐pentanedione in workplace air in a concentration range of a tenth up to twice the currently valid MAK value of 0.083 mg/m3. Sampling is performed by drawing a defined volume of air through a cartridge filled with silica gel impregnated with 2,4‐dinitrophenylhydrazine (DNPH) using a suitable flow‐regulated pump with a volumetric flow rate of 1 L/min. For sampling 1 hour or 15 min (checking the short‐term value) can be used. The formed hydrazones are desorbed with acetonitrile containing a defined amount of DNPH and analysed by means of liquid chromatography (HPLC‐UV or HPLC‐MS). Quantitative evaluation is based on a 10‐point calibration using an internal standard for analysis with HPLC‐MS and an external standard for analysis via HPLC‐UV. The limit of quantification for 2,3‐pentanedione is 0.00078 mg/m3 based on an air sample volume of approx. 60 L or 0.0031 mg/m3 based on an air sample volume of approx. 15 L (for short‐term value). Joint Publication of the Analytical Subcommittee of the Chemistry Board of Experts of the Expert Committee Raw Materials and Chemical Industry of the German Social Accident Insurance and the working group “Air Analyses” of the Permanent Senate Commission of the DFG for the Investigation of Health Hazards of Chemical Compounds in the Work Area.

Schlagwörter

LuftanalysenAnalysenmethodeArbeitsplatzmessungGefahrstoffBestimmungsgrenzeMessbereichProbenamePräzisionWiederfindungSelektivität

Zitieren

APA

Schuh, Claudia, Nitschke, Lutz, Hebisch, Ralph, Brock, Helmut, T., Hartwig, Andrea, & Commission, M. (2025). 2,3‐Pentanedione – Method for the determination of 2,3‐pentanedione in workplace air using high performance liquid chromatography (HPLC). The MAK Collection for Occupational Health and Safety, 3(3). https://doi.org/10.34865/am60014e1918a_w

Vancouver

Schuh, Claudia, Nitschke, Lutz, Hebisch, Ralph, et al. 2,3‐Pentanedione – Method for the determination of 2,3‐pentanedione in workplace air using high performance liquid chromatography (HPLC). The MAK Collection for Occupational Health and Safety. 2025;3(3). doi:10.34865/am60014e1918a_w

BibTeX
@article{Schuh2025,
  author = {Schuh and Claudia and Nitschke and Lutz and Hebisch and Ralph and Brock and Helmut, Thomas and Hartwig and Andrea and Commission, MAK},
  title = {2,3‐Pentanedione – Method for the determination of 2,3‐pentanedione in workplace air using high performance liquid chromatography (HPLC)},
  journal = {The MAK Collection for Occupational Health and Safety},
  year = {2025},
  volume = {3},
  number = {3},
  doi = {10.34865/am60014e1918a_w}
}
Kostenlos starten →