Motorized Leg Length Discrepancy Measure: A New Device for Clinical Use – A Cross-sectional Study
Crescent Journal of Medical and Biological Sciences · 2022
DOI: 10.34172/cjmb.2022.27 ↗Kurz gefasst
Querschnittstudie in der Zeitschrift Crescent Journal of Medical and Biological Sciences, 2022. Verfasst von Yaghoub Salekzamani und 4 weiteren Autor:innen. Die Arbeit wurde bisher 3-mal zitiert (Crossref).
Abstract (Originalsprache)
Objectives
This study aimed to construct a device that could measure leg length discrepancy (LLD) automatically. Materials and Methods: The LLD measure device measures LLD with pelvic-tilt method (Program 1) and weight-based method (Programs 2 & 3). Tests were done in 3 phases. 1: Two examiners using the LLD Measure device made -50 to 75 mm artificial LLD in two healthy subjects measuring the degree of pelvic tilt and the load bearing of lower limbs. 2: Sixteen healthy volunteers were asked to stand on the device to measure LLD with program 2 and then with both knees extended to measure LLD with program one. 3: 32 patients who had underwent lower limbs CT scanogram enrolled, and the LLD measurement with program 1 compared with those obtained by CT scanogram.
Results
Data’s obtained in the first phase showed excellent repeatability (intra-class correlation coefficient [ICC] > 0.9) and very good reproducibility (ICC > 80%) except for measuring the limb load while both knees were extended (ICC ≈ 60%). In the second phase, we found no statistically significant difference between measuring LLD using programs 1 and 2 (P = 0.49). In the third phase, there was no statistically significant difference between measuring LLD using program 1 and CT scanogram (P = 0.80).
Conclusions
We have developed a device to measure LLD semiautomatic with less need for examiner expertise. The results of our new device would be reliable and accurate compared to CT measurements.
Zitieren
APA
Salekzamani, Y., Fakhree, N. A., Ebrahimi, A., Heravi, H., & Dolatkhah, N. (2022). Motorized Leg Length Discrepancy Measure: A New Device for Clinical Use – A Cross-sectional Study. Crescent Journal of Medical and Biological Sciences. https://doi.org/10.34172/cjmb.2022.27
Vancouver
Salekzamani Y, Fakhree NA, Ebrahimi A, Heravi H, Dolatkhah N. Motorized Leg Length Discrepancy Measure: A New Device for Clinical Use – A Cross-sectional Study. Crescent Journal of Medical and Biological Sciences. 2022. doi:10.34172/cjmb.2022.27
BibTeX
@article{Salekzamani2022,
author = {Salekzamani, Yaghoub and Fakhree, Nargess Abolghassemi and Ebrahimi, Afshin and Heravi, Hamed and Dolatkhah, Neda},
title = {Motorized Leg Length Discrepancy Measure: A New Device for Clinical Use – A Cross-sectional Study},
journal = {Crescent Journal of Medical and Biological Sciences},
year = {2022},
doi = {10.34172/cjmb.2022.27}
}
EUVIMED