Computational analysis on design and optimization of low-frequency dielectrophoretic microfluidic lab on a chip for Staphylococcus aureus isolation from blood
BioImpacts · 13.06.2026 · 16 · 31358
DOI: 10.34172/bi.31358 ↗Kurz gefasst
Fachartikel in der Zeitschrift BioImpacts, online erschienen am 13.06.2026. Verfasst von Saeed Saedy und 3 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Engineering, Shahid Chamran University of Ahvaz, Ahvaz, Iran. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Introduction
Bloodstream infections (BSIs) caused by Staphylococcus aureus, particularly methicillin-resistant (MRSA) and methicillin-sensitive (MSSA) strains, remain a critical clinical concern due to high mortality rates and delays in diagnosis. Traditional approaches, such as blood culture, are time consuming, whereas molecular methods are limited by their inability to assess bacterial viability. Dielectrophoresis (DEP)-based lab-on-a-chip (LOC) technology offers a label-free, rapid, and viability preserving alternative for pathogen isolation by exploiting differences in dielectric properties.
Methods
A dual-channel microfluidic separation device was developed with independently operated pathways. The primary channel (2000×198 μm) integrated ten crown-shaped microelectrodes, while the secondary channel (1540×80 μm) incorporated eight microelectrodes. The primary channel isolated both target strains directly from whole blood; the secondary channel enabled further separation between standard Gram-positive MRSA and MSSA. Finite element modeling was performed using COMSOL Multiphysics, which integrated the Electric Currents, Creeping Flow, and Particle Tracing modules. Separation efficiency was evaluated across varying voltages, electrode geometries, and sample-to-buffer flow rate ratios at 1 Hz. Dielectric properties from published data were used to calculate the Clausius–Mossotti (CM) factor for predicting DEP behavior.
Results
In the primary channel, nearly 100% separation efficiency and purity were achieved for isolating MRSA and MSSA from red blood cells (RBCs), white blood cells (WBCs), and platelets (PLTs) at 24–28 Vpp and a sample-to-buffer flow rate ratio of 1:2. The secondary channel successfully differentiated MSSA from MRSA at 42–46 Vpp and a 1:5 flow rate ratio, exploiting their distinct membrane characteristics. Crown-shaped electrodes outperformed rectangular designs by generating higher electric field gradients, thereby enhancing separation performance.
Conclusion
The dual-channel LOC system enables high-efficiency, label-free separation and identification of MRSA and MSSA directly from whole blood using low-frequency negative DEP (nDEP). The optimized configuration supports rapid pathogen detection, offering substantial advantages over conventional diagnostic platforms.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Engineering, Shahid Chamran University of Ahvaz, Ahvaz, Iran.
ORCID-Profile: Saeed Saedy, Navid Alaei-Sheini
Zitieren
APA
Saedy, S., Alaei-Sheini, N., Ajabi, S., & Ansari-Asl, K. (2026). Computational analysis on design and optimization of low-frequency dielectrophoretic microfluidic lab on a chip for Staphylococcus aureus isolation from blood. BioImpacts, 16, 31358. https://doi.org/10.34172/bi.31358
Vancouver
Saedy S, Alaei-Sheini N, Ajabi S, Ansari-Asl K. Computational analysis on design and optimization of low-frequency dielectrophoretic microfluidic lab on a chip for Staphylococcus aureus isolation from blood. BioImpacts. 2026;16:31358. doi:10.34172/bi.31358
BibTeX
@article{Saedy2026,
author = {Saedy, Saeed and Alaei-Sheini, Navid and Ajabi, Shahrzad and Ansari-Asl, Karim},
title = {Computational analysis on design and optimization of low-frequency dielectrophoretic microfluidic lab on a chip for Staphylococcus aureus isolation from blood},
journal = {BioImpacts},
year = {2026},
volume = {16},
pages = {31358},
doi = {10.34172/bi.31358}
}
EUVIMED