Mechanical and Hemodynamic Effects of Balloon-Expandable and Self-Expanding Stents in a Saccular Abdominal Aortic Aneurysm Model: Fluid–Structure Interaction Analyses
Annals of Vascular Diseases · 2026 · 19 (1) · n/a
DOI: 10.3400/avd.oa.26-00052 ↗Kurz gefasst
Fachartikel in der Zeitschrift Annals of Vascular Diseases, 2026. Verfasst von Masanori Murakami und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Surgery, National Hospital Organization Kanmon Medical Center, Shimonoseki, Yamaguchi, Japan. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Japan Society for the Promotion of Science.
Abstract (Originalsprache)
Objectives
The aim of this study was to compare the effects of balloon-expandable and self-expanding stents on vessel wall mechanics and hemodynamics in a patient-specific saccular abdominal aortic aneurysm model.
Methods
A patient-specific aneurysm model was constructed using computed tomography images, and balloon-expandable and self-expanding stent models were created. Aortic wall stress was evaluated by contact analysis. Fluid-structure interaction analyses were performed at 5 representative phases selected from the cardiac cycle. Mean intra-aneurysmal flow velocity, wall shear stress (WSS), pressure loss, and energy loss were compared.
Results
The balloon-expandable stent produced greater localized aortic wall stress than the self-expanding stent, whereas wall stress was more uniformly distributed with the self-expanding stent. Both stents reduced the mean intra-aneurysmal flow velocity, WSS, pressure loss, and energy loss compared with the no-stent model. However, greater reductions in WSS, pressure loss, and energy loss were noted with the balloon-expandable stent than with the self-expanding stent.
Conclusions
Balloon-expandable and self-expanding stents exhibited different effects on vessel wall mechanics and hemodynamics. The findings indicate the importance of evaluating both hemodynamic and mechanical effects when assessing stent-vessel interactions.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiovascular Surgery, National Hospital Organization Kanmon Medical Center, Shimonoseki, Yamaguchi, Japan.
- Department of Mechanical Engineering, Graduate School of Sciences and Technology for Innovation, Yamaguchi University, Ube, Yamaguchi, Japan.
Förderung
- Japan Society for the Promotion of Science
Zitieren
APA
Murakami, M., Jiang, F., Yamashita, T., & Mochizuki, S. (2026). Mechanical and Hemodynamic Effects of Balloon-Expandable and Self-Expanding Stents in a Saccular Abdominal Aortic Aneurysm Model: Fluid–Structure Interaction Analyses. Annals of Vascular Diseases, 19(1), n/a. https://doi.org/10.3400/avd.oa.26-00052
Vancouver
Murakami M, Jiang F, Yamashita T, Mochizuki S. Mechanical and Hemodynamic Effects of Balloon-Expandable and Self-Expanding Stents in a Saccular Abdominal Aortic Aneurysm Model: Fluid–Structure Interaction Analyses. Annals of Vascular Diseases. 2026;19(1):n/a. doi:10.3400/avd.oa.26-00052
BibTeX
@article{Murakami2026,
author = {Murakami, Masanori and Jiang, Fei and Yamashita, Tatsuo and Mochizuki, Shinsuke},
title = {Mechanical and Hemodynamic Effects of Balloon-Expandable and Self-Expanding Stents in a Saccular Abdominal Aortic Aneurysm Model: Fluid–Structure Interaction Analyses},
journal = {Annals of Vascular Diseases},
year = {2026},
volume = {19},
number = {1},
pages = {n/a},
doi = {10.3400/avd.oa.26-00052}
}