EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Optimization and Validation of Dispersive Liquid–Liquid Microextraction for Simultaneous Determination of Aflatoxins B1, B2, G1, and G2 in Senna Leaves and Pods Using HPLC-FLD with Pre-Column Derivatization

Thanapoom Maneeboon iD, Chananya Chuaysrinule iD, Warapa Mahakarnchanakul

Toxins · 2023

DOI: 10.3390/toxins15040277 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Toxins, 2023. Verfasst von Thanapoom Maneeboon und 2 weiteren Autor:innen. Beteiligte Einrichtung: Scientific Equipment and Research Division, Kasetsart University Research and Development Institute (KURDI), Kasetsart University, Bangkok 10900, Tha…. Hauptthemen laut MEDLINE: Liquid Phase Microextraction, Aflatoxins. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Kasetsart University Research and Development Institute (KURDI), Kasetsart University, Bangkok, Thailand.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Dispersive liquid–liquid microextraction (DLLME) was optimized for the simultaneous extraction of aflatoxins (AFB1, AFB2, AFG1, and AFG2) from powdered senna leaves and pods. Detection was performed using high-performance liquid chromatography with fluorescence detection (HPLC-FLD) and pre-column derivatization. The parameters affecting the DLLME extraction efficiency were evaluated. Chloroform (200 µL) was used as an extraction solvent, 500 µL of distilled water was used as a dispersive solvent, and the extraction was performed at pH 5.6 with no salt added. The optimized method was validated using leaves and pods according to the European Commission guidelines. The linear range for all aflatoxins was 2–50 µg/kg, with values for regression coefficients of determination exceeding 0.995. The recoveries of spiked senna leaves and pods were in the ranges of 91.77–108.71% and 83.50–102.73%, respectively. The RSD values for intra-day and inter-day precisions were in the ranges of 2.30–7.93% and 3.13–10.59%, respectively. The limits of detection and quantification varied in the ranges of 0.70–1.27 µg/kg and 2.13–3.84 µg/kg, respectively. The validated method was successfully applied for the quantification of aflatoxins in 60 real samples of dried senna leaves and pods.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Liquid Phase MicroextractionAflatoxinsAflatoxin B1Chromatography, High Pressure LiquidLimit of Detection

Stichwörter der Autor:innen

DLLMEHPLC-FLDaflatoxinsmedicinal herbmycotoxinsenna

Substanzen

Aflatoxin B1Aflatoxins

Autor:innen und Institutionen

  • Scientific Equipment and Research Division, Kasetsart University Research and Development Institute (KURDI), Kasetsart University, Bangkok 10900, Thailand.
  • Department of Food Science and Technology, Faculty of Agro-Industry, Kasetsart University, Bangkok 10900, Thailand.

ORCID-Profile: Thanapoom Maneeboon, Chananya Chuaysrinule

Förderung

  • Kasetsart University Research and Development Institute (KURDI), Kasetsart University, Bangkok, Thailand

Zitieren

APA

Maneeboon, T., Chuaysrinule, C., & Mahakarnchanakul, W. (2023). Optimization and Validation of Dispersive Liquid–Liquid Microextraction for Simultaneous Determination of Aflatoxins B1, B2, G1, and G2 in Senna Leaves and Pods Using HPLC-FLD with Pre-Column Derivatization. Toxins. https://doi.org/10.3390/toxins15040277

Vancouver

Maneeboon T, Chuaysrinule C, Mahakarnchanakul W. Optimization and Validation of Dispersive Liquid–Liquid Microextraction for Simultaneous Determination of Aflatoxins B1, B2, G1, and G2 in Senna Leaves and Pods Using HPLC-FLD with Pre-Column Derivatization. Toxins. 2023. doi:10.3390/toxins15040277

BibTeX
@article{Maneeboon2023,
  author = {Maneeboon, Thanapoom and Chuaysrinule, Chananya and Mahakarnchanakul, Warapa},
  title = {Optimization and Validation of Dispersive Liquid–Liquid Microextraction for Simultaneous Determination of Aflatoxins B1, B2, G1, and G2 in Senna Leaves and Pods Using HPLC-FLD with Pre-Column Derivatization},
  journal = {Toxins},
  year = {2023},
  doi = {10.3390/toxins15040277}
}
Kostenlos starten →