Where Does the Signal Go? Technical Challenges of CT Perfusion in Lacunar and Infratentorial Stroke
Tomography · 21.08.2026 · 12 (8) · 118
DOI: 10.3390/tomography12080118 ↗Kurz gefasst
Übersichtsarbeit in der Zeitschrift Tomography, online erschienen am 21.08.2026. Verfasst von Nicola Morelli und 4 weiteren Autor:innen. Beteiligte Einrichtung: Diagnostic Neuroradiology Unit, Guglielmo da Saliceto Hospital, Azienda USL di Piacenza, 29121 Piacenza, Italy. Hauptthemen laut MEDLINE: Stroke, Lacunar, Tomography, X-Ray Computed, Stroke. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 59 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Schlaganfall
Abstract (Originalsprache)
Background/Objectives: Computed tomography perfusion is widely used in acute ischemic stroke, but its performance is less reliable for lacunar and infratentorial infarcts. This technical review examines the acquisition and processing factors that influence their detectability.
Methods
A targeted technical review of PubMed/MEDLINE, Scopus, and Web of Science was performed, focusing on acquisition, reconstruction, vascular input selection, deconvolution, filtering, spatial sampling, automated classification, posterior circulation stroke, and lacunar infarction.
Results
Detectability depends on lesion size and contrast, posterior fossa artifacts, spatial and temporal resolution, vascular curve quality, mathematical stabilization, and automated thresholds. Larger cerebellar infarcts may remain visible, whereas brainstem, deep cerebellar, and perforator lesions are more vulnerable. Temporal maps and direct review of parametric images may reveal abnormalities absent from automated summaries.
Conclusions
This technical review shows that computed tomography perfusion should be interpreted as a derived estimate rather than a direct representation of cerebral hemodynamics. Negative automated findings do not exclude lacunar or infratentorial infarction when clinical suspicion remains high.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Diagnostic Neuroradiology Unit, Guglielmo da Saliceto Hospital, Azienda USL di Piacenza, 29121 Piacenza, Italy.
- Neurology Unit, Department of Medicine and Surgery, University of Parma, 43126 Parma, Italy.
- Neurology Unit, San Giacomo Hospital, 15067 Novi Ligure, Italy.
- Department of Diagnostic Imaging, Centro Diagnostico Rocca, 29121 Piacenza, Italy.
ORCID-Profile: Nicola Morelli, Marco Spallazzi, Eugenia Rota, Davide Colombi
Zitieren
APA
Morelli, N., Spallazzi, M., Rota, E., Biondi, M., & Colombi, D. (2026). Where Does the Signal Go? Technical Challenges of CT Perfusion in Lacunar and Infratentorial Stroke. Tomography, 12(8), 118. https://doi.org/10.3390/tomography12080118
Vancouver
Morelli N, Spallazzi M, Rota E, Biondi M, Colombi D. Where Does the Signal Go? Technical Challenges of CT Perfusion in Lacunar and Infratentorial Stroke. Tomography. 2026;12(8):118. doi:10.3390/tomography12080118
BibTeX
@article{Morelli2026,
author = {Morelli, Nicola and Spallazzi, Marco and Rota, Eugenia and Biondi, Marina and Colombi, Davide},
title = {Where Does the Signal Go? Technical Challenges of CT Perfusion in Lacunar and Infratentorial Stroke},
journal = {Tomography},
year = {2026},
volume = {12},
number = {8},
pages = {118},
doi = {10.3390/tomography12080118}
}