AI-Driven Prediction of Chest CT Radiation Doses: Establishing BMI-Based Diagnostic Reference Levels and Patient–Factor Correlations for Machine-Learning Models
Tomography · 28.04.2026 · 12 (5) · 61
DOI: 10.3390/tomography12050061 ↗Kurz gefasst
Fachartikel in der Zeitschrift Tomography, online erschienen am 28.04.2026. Verfasst von Zuhal Y Hamd und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Sciences, College of Health and Rehabilitation Sciences, Princess Nourah bint Abdulrahman University, P.O. Box 84428, Riya…. Hauptthemen laut MEDLINE: Machine Learning, Body Mass Index, Radiation Dosage. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deanship of Scientific Research and Libraries at Princess Nourah bint Abdulrahman University, through the, Deanship of Scientific Research and Libraries.
Abstract (Originalsprache)
Background and aim: Chest CT is a major contributor to population radiation exposure. Conventional, pooled diagnostic reference levels (DRLs) do not account for inter-individual variability in body habitus and are typically used retrospectively. We evaluated dose behavior in adult chest CT, derived BMI-stratified local DRLs, and developed models to enable AI-assisted, prescan dose prediction.
Methods
Consecutive adult chest CT examinations from a single center were analyzed. Dose indices (CTDIvol, DLP) and patient factors (BMI, weight, height, age, sex; scan length and planned technical parameters where available) were extracted. DRLs were defined as the 75th percentile overall and within BMI categories (underweight, normal, overweight, and obese). Group differences were assessed using non-parametric tests; associations were examined using correlation analysis. Supervised learning (e.g., Random Forest, Gradient Boosting) was trained to predict CTDIvol and DLP from routinely available variables.
Results
BMI-stratified DRLs increased monotonically with habitus: underweight 444.95 mGy·cm/9.60 mGy; normal 513.00/11.55; overweight 756.08/14.65; obese 931.60/20.25 (DLP/CTDIvol). Differences across BMI groups were significant for DLP (H = 31.53, p < 0.001) and CTDIvol (H = 33.61, p < 0.001). DLP correlated moderately with weight and BMI (r ≈ 0.54–0.56, p < 0.001), with a weaker association for age; height was not a meaningful predictor. No sex-based differences in CTDIvol or DLP were observed. Predictive models estimated CTDIvol and DLP with high performance (R2 up to ~0.79 and ~0.77, respectively), enabling comparison of predicted dose against BMI-matched DRLs before acquisition.
Conclusions
Size-aware, BMI-stratified DRLs provide clinically interpretable investigation levels that avoid pitfalls of pooled benchmarks. Coupled with robust prediction of individualized dose from routine variables, this framework supports a shift from retrospective audit to prospective, point-of-care dose governance and protocol optimization in chest CT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiological Sciences, College of Health and Rehabilitation Sciences, Princess Nourah bint Abdulrahman University, P.O. Box 84428, Riyadh 11671, Saudi Arabia.
- Medical Diagnostic Imaging Department, College of Health Sciences, University of Sharjah, Sharjah P.O. Box 26666, United Arab Emirates.
- Medical Imaging Department, King Abdullah bin Abdul-Aziz University Hospital, Riyadh 11564, Saudi Arabia.
- Department of Physics, College of Sciences, Princess Nourah bint Abdulrahman University, P.O. Box 84428, Riyadh 11671, Saudi Arabia.
- Altakassusi Alliance Medical LLC, King Fahad Medical City, Riyadh 12311, Saudi Arabia.
ORCID-Profile: Zuhal Y Hamd, Mohamed Abuzaid, Nissren Tamam, Amal I Alorainy
Förderung
- Deanship of Scientific Research and Libraries at Princess Nourah bint Abdulrahman University, through the
- Deanship of Scientific Research and Libraries
Zitieren
APA
Hamd, Z. Y., Abuzaid, M., Alharbi, M., Tamam, N., Alorainy, A. I., Alrujaee, L., Almutairi, N., & Alyagoub, A. A. (2026). AI-Driven Prediction of Chest CT Radiation Doses: Establishing BMI-Based Diagnostic Reference Levels and Patient–Factor Correlations for Machine-Learning Models. Tomography, 12(5), 61. https://doi.org/10.3390/tomography12050061
Vancouver
Hamd ZY, Abuzaid M, Alharbi M, Tamam N, Alorainy AI, Alrujaee L, et al. AI-Driven Prediction of Chest CT Radiation Doses: Establishing BMI-Based Diagnostic Reference Levels and Patient–Factor Correlations for Machine-Learning Models. Tomography. 2026;12(5):61. doi:10.3390/tomography12050061
BibTeX
@article{Hamd2026,
author = {Hamd, Zuhal Y and Abuzaid, Mohamed and Alharbi, Mohamed and Tamam, Nissren and Alorainy, Amal I and Alrujaee, Lena and Almutairi, Najla and Alyagoub, Aljouharah Abdullah},
title = {AI-Driven Prediction of Chest CT Radiation Doses: Establishing BMI-Based Diagnostic Reference Levels and Patient–Factor Correlations for Machine-Learning Models},
journal = {Tomography},
year = {2026},
volume = {12},
number = {5},
pages = {61},
doi = {10.3390/tomography12050061}
}
EUVIMED