Optimizing Microbubble Reduction to Facilitate IVUS Guidance During Endovascular Radiofrequency Wire Procedures
Tomography · 31.03.2026 · 12 (4) · 48
DOI: 10.3390/tomography12040048 ↗Kurz gefasst
Fachartikel in der Zeitschrift Tomography, online erschienen am 31.03.2026. Verfasst von Curtis Plante und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Interventional Radiology, Department of Radiology, University of Vermont Medical Center, 111 Colchester Ave., Burlington, VT 05401, USA. Hauptthemen laut MEDLINE: Microbubbles, Ultrasonography, Interventional, Liver, Endovascular Procedures. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: Radiofrequency (RF) wire energy can be used for tissue ablation across many conditions. Adjusting RF generator parameters allows RF energy to puncture tissue with minimal adjacent damage. When RF energy is applied to tissue, however, microbubbles are produced, obstructing intravascular ultrasound (IVUS). Mitigation of RF-generated microbubbles has been studied for ablation but not for puncture.
Methods
This paper describes an in vitro bench study using ex vivo bovine live tissue. A model was created with bovine liver tissue and an IVUS catheter submerged in a saline bath. Tissue was punctured with an RF guidewire to recreate microbubbles. Following the puncture, various methods were applied: altering the mechanical index of the IVUS, applying a VF10-5 Linear probe (Siemens), and applying a L12-3 Linear probe (Philips). Regions of interest (ROIs) were selected to track pixel brightness as a proxy for microbubbles.
Results
The control increased ROI brightness by 1.5%. Altering the mechanical index of IVUS reduced ROI brightness by 1.2%. VF10-5 probe application increased ROI brightness by 1.2%. L12-3 probe application reduced ROI brightness by 33.0% (p = 0.046, n = 3, one-sample t-test). Brightness reduction was most pronounced at the site of initial RF wire puncture, where microbubbles accumulated. Tip visualization improved, allowing for more precise wire trajectory adjustments.
Conclusions
External US with an L12-3 probe was able to dissipate microbubbles effectively to improve IVUS guidance following RF wire puncture in an in vitro exploratory bench model.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Interventional Radiology, Department of Radiology, University of Vermont Medical Center, 111 Colchester Ave., Burlington, VT 05401, USA.
- Department of Chemical Engineering, Queen's University, 19 Division St, Kingston, ON K7L 2N9, Canada.
- Department of Medicine, Queen's University, 94 Stuart Street, Kingston, ON K7L 3N6, Canada.
- Department of Cardiovascular Diseases, Mayo Clinic, Rochester, MN 55905, USA.
ORCID-Profile: Curtis Plante, Carlos Escobedo, David S Majdalany, Bill S Majdalany
Zitieren
APA
Plante, C., Warfield, A. E., Escobedo, C., Johri, A. M., Majdalany, D. S., & Majdalany, B. S. (2026). Optimizing Microbubble Reduction to Facilitate IVUS Guidance During Endovascular Radiofrequency Wire Procedures. Tomography, 12(4), 48. https://doi.org/10.3390/tomography12040048
Vancouver
Plante C, Warfield AE, Escobedo C, Johri AM, Majdalany DS, Majdalany BS. Optimizing Microbubble Reduction to Facilitate IVUS Guidance During Endovascular Radiofrequency Wire Procedures. Tomography. 2026;12(4):48. doi:10.3390/tomography12040048
BibTeX
@article{Plante2026,
author = {Plante, Curtis and Warfield, Andrew E and Escobedo, Carlos and Johri, Amer M and Majdalany, David S and Majdalany, Bill S},
title = {Optimizing Microbubble Reduction to Facilitate IVUS Guidance During Endovascular Radiofrequency Wire Procedures},
journal = {Tomography},
year = {2026},
volume = {12},
number = {4},
pages = {48},
doi = {10.3390/tomography12040048}
}
EUVIMED