Optimizing Imaging Parameters for Assessment of Hepatocellular Carcinoma Using Photon-Counting Detector Computed Tomography—Impact of Reconstruction Kernel and Slice Thickness

Anna Szelenyi, Philipp Stelzer, Christian Wassipaul iD, Jakob Kittinger, Andreas Strassl iD, Victor Schmidbauer iD, Martin Luther Watzenböck, Florian Lindenlaub iD, Michael Arnoldner, Michael Weber iD, Matthias Pinter iD, Ruxandra-Iulia Milos iD und 1 weitere

Tomography · 27.06.2025 · 11 (7) · 77

DOI: 10.3390/tomography11070077 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Tomography, online erschienen am 27.06.2025. Verfasst von Anna Szelenyi und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, 1090 Vienna, Austria. Hauptthemen laut MEDLINE: Carcinoma, Hepatocellular, Liver Neoplasms, Tomography, X-Ray Computed, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Siemens Healthineers AG.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The use of photon-counting detector computed tomography (PCD-CT) has improved image quality in cardiac, pulmonary, and musculoskeletal imaging. Abdominal imaging research, especially about the use of PCD-CT in hepatocellular carcinoma (HCC), is sparse.

Objectives

We aimed to compare the image quality of tumors, the liver parenchyma, and the vasculature in patients with HCC using PCD-CT reconstructions at different slice thicknesses and kernels to identify the most appropriate settings for the clinical routine.

Methods

CT exams from twenty adult patients with HCC performed with a clinically approved, first-generation PCD-CT scanner (Naeotom Alpha®, Siemens Healthineers), were retrospectively reviewed. For each patient, images were reconstructed at four different sharp kernels, designed for abdominal imaging (Br40; Br44; Br48; Br56) and at three slice thicknesses (0.4 mm; 1 mm; 3 mm). The reconstruction with the Br40 kernel at 3 mm (Br403 mm) was used as a clinical reference. Three readers independently assessed the image quality of different anatomical abdominal structures and hypervascular HCC lesions using a five-point Likert scale. In addition, image sharpness was assessed using line-density profiles.

Results

Compared with the clinical reference, the Br441 mm and Br481 mm reconstructions were rated superior for the assessment of the hepatic vasculature (median difference +0.67 [+0.33 to +1.33], p < 0.001 and +1.00 [+0.67 to +1.67], p < 0.001). Reconstructions for Br401 mm (+0.33 [−0.67 to +1.00], p < 0.001), and Br443 mm (+0.0 [0.0 to +1.00], p = 0.030) were scored superior for overall image quality. The noise demonstrated a continuous increase when using sharper kernels and thinner slices than Br403 mm (p < 0.001), leading to a decrease in contrast-to-noise ratio. Although there was a trend toward increased image sharpness using the slope analysis with higher kernels, this was not significantly different compared with the reference standard.

Conclusion

PCD-CT reconstruction Br401 mm was the most suitable setting for overall image quality, while reconstructions with sharper kernels (Br441 mm and Br481 mm) can be considered for the assessment of the hepatic vasculature in patients with HCC.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Carcinoma, HepatocellularLiver NeoplasmsTomography, X-Ray ComputedRadiographic Image Interpretation, Computer-AssistedHumansMaleFemaleMiddle AgedRetrospective StudiesAgedAdultLiver

Stichwörter der Autor:innen

hepatocellular carcinomakernel reconstructionsphoton counting detector computed tomographyspatial resolution

Autor:innen und Institutionen

  • Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, 1090 Vienna, Austria.
  • Division of Gastroenterology and Hepatology, Medical University of Vienna, 1090 Vienna, Austria.

ORCID-Profile: Christian Wassipaul, Andreas Strassl, Victor Schmidbauer, Florian Lindenlaub, Michael Weber, Matthias Pinter, Ruxandra-Iulia Milos

Förderung

  • Siemens Healthineers AG

Zitieren

APA

Szelenyi, A., Stelzer, P., Wassipaul, C., Kittinger, J., Strassl, A., Schmidbauer, V., Watzenböck, M. L., Lindenlaub, F., Arnoldner, M., Weber, M., Pinter, M., Milos, R. I., & Tamandl, D. (2025). Optimizing Imaging Parameters for Assessment of Hepatocellular Carcinoma Using Photon-Counting Detector Computed Tomography—Impact of Reconstruction Kernel and Slice Thickness. Tomography, 11(7), 77. https://doi.org/10.3390/tomography11070077

Vancouver

Szelenyi A, Stelzer P, Wassipaul C, Kittinger J, Strassl A, Schmidbauer V, et al. Optimizing Imaging Parameters for Assessment of Hepatocellular Carcinoma Using Photon-Counting Detector Computed Tomography—Impact of Reconstruction Kernel and Slice Thickness. Tomography. 2025;11(7):77. doi:10.3390/tomography11070077

BibTeX
@article{Szelenyi2025,
  author = {Szelenyi, Anna and Stelzer, Philipp and Wassipaul, Christian and Kittinger, Jakob and Strassl, Andreas and Schmidbauer, Victor and Watzenböck, Martin Luther and Lindenlaub, Florian and Arnoldner, Michael and Weber, Michael and Pinter, Matthias and Milos, Ruxandra-Iulia and Tamandl, Dietmar},
  title = {Optimizing Imaging Parameters for Assessment of Hepatocellular Carcinoma Using Photon-Counting Detector Computed Tomography—Impact of Reconstruction Kernel and Slice Thickness},
  journal = {Tomography},
  year = {2025},
  volume = {11},
  number = {7},
  pages = {77},
  doi = {10.3390/tomography11070077}
}
Kostenlos starten →