Noise Reduction in Brain CT: A Comparative Study of Deep Learning and Hybrid Iterative Reconstruction Using Multiple Parameters
Tomography · 2024
DOI: 10.3390/tomography10120147 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Tomography, 2024. Verfasst von Yusuke Inoue und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic Radiology, Kitasato University School of Medicine, Sagamihara 252-0374, Japan. Hauptthemen laut MEDLINE: Deep Learning, Phantoms, Imaging, Tomography, X-Ray Computed, Brain. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
We evaluated the noise reduction effects of deep learning reconstruction (DLR) and hybrid iterative reconstruction (HIR) in brain computed tomography (CT).
Methods
CT images of a 16 cm dosimetry phantom, a head phantom, and the brains of 11 patients were reconstructed using filtered backprojection (FBP) and various levels of DLR and HIR. The slice thickness was 5, 2.5, 1.25, and 0.625 mm. Phantom imaging was also conducted at various tube currents. The noise reduction ratio was calculated using FBP as the reference. For patient imaging, overall image quality was visually compared between DLR and HIR images that exhibited similar noise reduction ratios.
Results
The noise reduction ratio increased with increasing levels of DLR and HIR in phantom and patient imaging. For DLR, noise reduction was more pronounced with decreasing slice thickness, while such thickness dependence was less evident for HIR. Although the noise reduction effects of DLR were similar between the head phantom and patients, they differed for the dosimetry phantom. Variations between imaging objects were small for HIR. The noise reduction ratio was low at low tube currents for the dosimetry phantom using DLR; otherwise, the influence of the tube current was small. In terms of visual image quality, DLR outperformed HIR in 1.25 mm thick images but not in thicker images.
Conclusions
The degree of noise reduction using DLR depends on the slice thickness, tube current, and imaging object in addition to the level of DLR, which should be considered in the clinical use of DLR. DLR may be particularly beneficial for thin-slice imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Diagnostic Radiology, Kitasato University School of Medicine, Sagamihara 252-0374, Japan.
- Department of Radiology, Kitasato University Hospital, Sagamihara 252-0375, Japan.
ORCID-Profile: Yusuke Inoue
Zitieren
APA
Inoue, Y., Itoh, H., Hata, H., Miyatake, H., Mitsui, K., Uehara, S., & Masuda, C. (2024). Noise Reduction in Brain CT: A Comparative Study of Deep Learning and Hybrid Iterative Reconstruction Using Multiple Parameters. Tomography. https://doi.org/10.3390/tomography10120147
Vancouver
Inoue Y, Itoh H, Hata H, Miyatake H, Mitsui K, Uehara S, et al. Noise Reduction in Brain CT: A Comparative Study of Deep Learning and Hybrid Iterative Reconstruction Using Multiple Parameters. Tomography. 2024. doi:10.3390/tomography10120147
BibTeX
@article{Inoue2024,
author = {Inoue, Yusuke and Itoh, Hiroyasu and Hata, Hirofumi and Miyatake, Hiroki and Mitsui, Kohei and Uehara, Shunichi and Masuda, Chisaki},
title = {Noise Reduction in Brain CT: A Comparative Study of Deep Learning and Hybrid Iterative Reconstruction Using Multiple Parameters},
journal = {Tomography},
year = {2024},
doi = {10.3390/tomography10120147}
}
EUVIMED